Dr. Sahbi Aloui and Dr. Horst DeckmannPublished on 01.11.2020 # Influence of the Heating Rate on the Glass Transition Temperature of Elastomer Materials for the DMA GABO *Eplexor*® Series - DMA EPLEXOR 500N - Polymers #### Introduction Elastomer materials are employed in almost all technical fields due to their high elasticity. An essential property of elastomer materials is the capability of storing deformation energy and releasing it back to the entire system, when necessary. One measure of this property consists of the material-immanent restoring forces, which – depending on the system – can be generated from the energy stored and can easily amount to 90% or more of the energy stored. This “valuable“ property is restricted, however, to a narrow temperature range that defines the operating and working temperatures for the respective application. For this reason, the temperature behavior of elastomer materials is of central importance. So-called temperature sweeps are used to record the thermal behavior of elastomer materials, which can generally be parameterized at different heating rates. A high heating rate of 5°C/min, for example, is preferable to a heating rate of 1°C/min since a result is delivered in a shorter time and testing is therefore faster and more cost-effective. However, the question arises as to how to evaluate the results for different heating rates. This Application Note addresses this question and examines the heating-rate dependence of the DMA GABO *Eplexor*® series. **Figure:** 1) DMA GABO EPLEXOR 500 N ![1) DMA GABO EPLEXOR 500 N](https://analyzing-testing.netzsch.com/_Resources/Persistent/e/d/e/3/ede3279377f258e4962c08d20689e01c5d3246c9/NETZSCH_AN_201_Abb_1.png) The photograph shows a DMA GABO EPLEXOR 500 N viscometer on a laboratory bench, with its test chamber, vertical loading assembly and control unit visible. A computer monitor beside the instrument displays measurement data for material analysis. #### Measurement Conditions Four temperature sweeps on samples of the same rubber compound were carried out from -80°C to 20°C at heating rates of 1, 2, 3 and 5°C/min with the DMA GABO *Eplexor*® 500 N (figure 1). #### Introduction The lower operating temperature of elastomer materials is limited by the glass transition temperature, Tg. The Tg characterizes the temperature at which elastomer materials change from a hard and relatively brittle state to a rubber-like elastic state. In practice, the Tg is defined as the maximum of the loss factor tanδ. The heating-rate dependence of the Tg is depicted in figure 1. Figure 2 shows that the Tg is shifted to higher temperatures with higher heating rates. For a temperature sweep, the Tg amounts to -42.3°C at a heating rate of 1 °C/min and to -41.4°C at a heating rate of 5 °C/min. This corresponds to a positional change of the Tg of approx. 1°C. The maximum of the loss factor, tanδ, has changed by 0.01 at most. This observation can be illustrated by the poor thermal conductivity of most plastics. This causes a shift in material-specific transition effects, such as relaxation maxima or glass transition temperatures, to higher temperatures (in the case of positive heating rates) or to lower temperatures (in the case of negative cooling rates). A higher heating rate leads to “drag effects” and the sample lags behind the furnace temperature. A heating rate of 1°C/min will therefore correctly reflect the sample-specific effects, while a high heating rate will cause a shift of these effects on the temperature scale. **Figure:** 2) Heating-rate dependency of the glass transition temperature, Tg ![2) Heating-rate dependency of the glass transition temperature, Tg](https://analyzing-testing.netzsch.com/_Resources/Persistent/f/a/f/9/faf905ca896c4c620c7937618a93cbd868ef17db/NETZSCH_AN_201_Abb_2.png) The chart shows DMA results for the heating-rate dependence of the glass transition temperature and maximum loss factor, tan δ. The x-axis is the heating rate in °C/min. Blue circles and the blue fitted line represent the glass transition temperature, Tg, on the left y-axis in °C; red squares represent the maximum tan δ on the right y-axis. At heating rates of 1, 2, 3 and 5 °C/min, Tg is approximately −42.3, −42.1, −41.6 and −41.5 °C, respectively. The maximum tan δ is approximately 0.973, 0.976, 0.973 and 0.966, respectively. Tg shifts slightly toward higher temperatures as the heating rate increases, while the maximum tan δ varies only slightly and is lower at 5 °C/min. The figure demonstrates the minimal heating-rate dependence of the glass transition position and tan δ maximum, consistent with the reported good temperature distribution in the DMA GABO EPLEXOR furnace. #### Summary These minimal shifts of the Tg’s position and the maximum of the loss factor, tanδ, as a result of the different heating rates are due to a very good temperature distribution inside the DMA GABO *Eplexor*® series, achieved by using a fan in the measuring chamber. A direct consequence of these findings is the reduction of the measuring time needed for temperature sweeps via the use of higher heating rates, for example, 5°C/min instead of, for example, 1°C/min. A prerequisite for this is knowledge of the heating-rate dependence of the Tg of the materials tested. Products [DMA *Eplexor*® up to 500 N](https://analyzing-testing.netzsch.com/en/footer/legacy-products/dma-eplexor-up-to-500-n.mdown) Content - Measurement Conditions - Introduction - Summary Related Application Notes[Download PDF](https://analyzing-testing.netzsch.com/_Resources/Persistent/6/7/6/2/6762d1e2186de3c2ee86c3e79b4aa2bb9807292c/AN201_~1.pdf) [Do you have any questions? Contact us here!](https://analyzing-testing.netzsch.com/en/Formularseiten/general-contact-request-form.mdown) #### Related Application Notes [PolymersRheologyKinexus Prime Series](https://analyzing-testing.netzsch.com/en/application-literature/chirps-measurements-on-lldpe-with-the-kinexus.mdown) A Broader Frequency Range in Less Time: Chirps Measurements on LLDPE with the KinexusAchieve broader frequency range and 80% faster rheological measurements on LLDPE with Kinexus chirps, enhancing polymer analysis efficiency and precision. [PolymersSTASTA 319 *Jupiter®*TGADSCPeakSeparation](https://analyzing-testing.netzsch.com/en/application-literature/reliable-assessment-of-recyclates-simultaneous-tg-dsc-analysis-of-pe-pp-blends.mdown) Reliable Assessment of Recyclates: Simultaneous TG-DSC Analysis of PE-PP Blends [PolymersLFALFA 717 *HyperFlash®*](https://analyzing-testing.netzsch.com/en/application-literature/thermal-diffusivity-of-3d-printed-pla-filaments-influence-of-the-print-orientation.mdown) Thermal Diffusivity of 3D-Printed PLA Filaments: Influence of the Print OrientationDiscover how print orientation affects thermal diffusivity in 3D-printed PLA filaments using NETZSCH LFA 717 *HyperFlash®* for precise analysis. [PolymersDSCDSC 300 *Caliris®* *Supreme*](https://analyzing-testing.netzsch.com/en/application-literature/visual-characterization-of-the-freeze-thaw-process-of-pva-solutions-by-with-integrated-dsc-camera.mdown) Visual Characterization of the Freeze-Thaw Process of PVA Solutions with Integrated DSC CameraVisualize and analyze the freeze-thaw process of PVA solutions with NETZSCH’s integrated DSC camera, revealing thermal and structural hydrogel changes. [PolymersRheologyCapillary Rheometer](https://analyzing-testing.netzsch.com/en/application-literature/freeflow-dies-no-sticking-no-measurement-errors-reliable-extensional-viscosity-data.mdown) FreeFlow Dies: No Sticking. No Measurement Errors. Reliable Extensional Viscosity Data [PolymersDEADEA 288 *Ionic*RheologyKinexus Prime Series](https://analyzing-testing.netzsch.com/en/application-literature/simultaneous-rheological-and-dielectric-characterization-of-an-amine-cured-epoxy-resin.mdown) Simultaneous Rheological and Dielectric Characterization of an Amine-Cured Epoxy ResinSimultaneously track mechanical and molecular changes during amine-cured epoxy resin curing using NETZSCH’s rheology and dielectric analysis tools. [PolymersRheologyRosand RH7/RH10](https://analyzing-testing.netzsch.com/en/application-literature/study-of-the-effect-of-moisture-on-the-viscosity-of-pa-using-capillary-rheometry.mdown) Study of the Effect of Moisture on the Viscosity of PA Using Capillary RheometryMonitor amine-cured epoxy cure with NETZSCH Kinexus Prime rheometer and DEA 288 dielectric analyzer for precise control. [PolymersDSCDSC 300 *Caliris®* *Supreme**Kinetics Neo*](https://analyzing-testing.netzsch.com/en/application-literature/crystallization-kinetics-of-polyamide-12-for-additive-manufacturing-processing.mdown) Crystallization Kinetics of Polyamide 12 for Additive Manufacturing Processing