BKrhSXOJ05eAh89d6e6ee0051abfeaebda5e3d07d5c67BKrhSXOJ05eAhNode\_%d0dbe915091d400bd8ee7f27f0791303%\_821b195c-4cf0-4bb6-bcf5-5e1ecc0b49df,DescendantOf\_%d0dbe915091d400bd8ee7f27f0791303%\_821b195c-4cf0-4bb6-bcf5-5e1ecc0b49dfBKrhSXOJ05eAh BKrhSXOJ05eAh005e2ebee53750c0b2649fbbf41c85a9BKrhSXOJ05eAhNodeType\_%d0dbe915091d400bd8ee7f27f0791303%\_Medienreaktor.Site:Document.Page.HomepageBKrhSXOJ05eAhBKrhSXOJ05eAhBKrhSXOJ05eAhe37849119d6f56e7aa1b969896984673BKrhSXOJ05eAhNodeType\_%d0dbe915091d400bd8ee7f27f0791303%\_Medienreaktor.Site:Document.Page.HomepageBKrhSXOJ05eAhBKrhSXOJ05eAhBKrhSXOJ05eAhf2dc8c645c09e35794185f5a379229f9BKrhSXOJ05eAhNodeType\_%d0dbe915091d400bd8ee7f27f0791303%\_Medienreaktor.Site:Document.Page.HomepageBKrhSXOJ05eAhBKrhSXOJ05eAh # Principle of the LFA Method The laser or light flash method dates back to studies by Parker et al. in 1961. In carrying out a measurement, the lower surface of a plane parallel sample (see fig. 1) is first heated by a short energy pulse. The resulting temperature change on the upper surface of the sample is then measured with an infrared detector. The typical course of the signals is presented in figure 2 (red curve). The higher the sample’s thermal diffusivity, the steeper the signal increase. *a*: Thermal diffusivity ρ: Density cp: Specific heat capacity λ: Thermal conductivity T: Temperature Using the half time (t1/2, time value at half signal height) and sample thickness (d), the thermal diffusivity (*a*) and finally the thermal conductivity (λ) can be calculated by means of the formula in figure 2. Furthermore, the specific heat (cp) of solids can be determined using the signal height (ΔTmax) compared to the signal height of a reference material. LFA investigations generally take much less time than thermal conductivity measurements by means of GHP (Guarded Hot Plate) or HFM (Heat Flow Meter). **Figure:** Diagram of the Laser Flash Analysis method, illustrating sample placement, heating elements, and detector positions for thermal conductivity measurement. ![Diagram of the Laser Flash Analysis method, illustrating sample placement, heating elements, and detector positions for thermal conductivity measurement.](https://analyzing-testing.netzsch.com/_Resources/Persistent/2/f/a/0/2fa04495a03633d048c7a855811ab56db9568457/LFA_467_HyperFlash_005-1414x1010.png) The diagram illustrates the flash technique used in Laser Flash Analysis (LFA). A disk-shaped “sample” is positioned centrally inside a vertical “protective tube,” with a “heating element/furnace” on the right and a “power source” below. Upward arrows indicate heat propagation through the sample toward the upper “detector,” which records the temperature response. A “sample thermo-couple” is shown extending from the left into the sample position. The arrangement depicts heating from below and temperature detection above for determining thermal transport properties. **Figure:** Graph illustrating the laser flash analysis method with temperature change over time, showcasing key formulas for thermal conductivity. ![Graph illustrating the laser flash analysis method with temperature change over time, showcasing key formulas for thermal conductivity.](https://analyzing-testing.netzsch.com/_Resources/Persistent/b/6/4/f/b64f9e48e04aeee9df029592d0da1492a915c2f5/csm_LFA_467_HyperFlash_Flash_Technique_2_e97dc6394b-500x357.png) The diagram shows the temperature increase measured in laser flash analysis (LFA) as a function of time. The vertical axis is the temperature change, ΔT, and the horizontal axis is time, t; no units are shown. A red sigmoidal curve rises from near zero to the maximum temperature change, ΔT_max. The half-maximum, 1/2 ΔT_max, defines the half-rise time, t_1/2. The thermal diffusivity is calculated as \(a = 0.1388\,d^2/t_{1/2}\), where d is the sample thickness. The maximum temperature change is shown as proportional to \(1/(m·c_p)\), with m denoting sample mass and \(c_p\) specific heat capacity. The lower formula gives thermal conductivity as \(\lambda(T)=a(T)·c_p(T)·\rho(T)\), with \(\rho(T)\) denoting density. The diagram illustrates that LFA determines thermal diffusivity from the time-dependent rear-face temperature response and that thermal conductivity is obtained from diffusivity, specific heat capacity and density. back to the products BKrhSXOJ05eAhae30ef1d873f4ac847201e903c6b541bBKrhSXOJ05eAhNodeType_%d0dbe915091d400bd8ee7f27f0791303%_Medienreaktor.Site:Document.Product,Node_%d0dbe915091d400bd8ee7f27f0791303%_bab15058-9ec6-4168-9a13-ea3eed259e89,Node_%d0dbe915091d400bd8ee7f27f0791303%_821b195c-4cf0-4bb6-bcf5-5e1ecc0b49dfBKrhSXOJ05eAh- [LFA 427](https://analyzing-testing.netzsch.com/en/products/thermal-diffusivity-and-conductivity/lfa-427.mdown) - [LFA 457 *MicroFlash®*](https://analyzing-testing.netzsch.com/en/footer/legacy-products/lfa-457-micro-flash.mdown) - [LFA 467 *HyperFlash®*](https://analyzing-testing.netzsch.com/en/footer/legacy-products/lfa-467-hyper-flash-light-flash-apparatus.mdown) BKrhSXOJ05eAh BKrhSXOJ05eAhd8bcd194b9a76ccd5b11198b0a7745a7BKrhSXOJ05eAhNodeType\_%d0dbe915091d400bd8ee7f27f0791303%\_Medienreaktor.Site:Document.Page.HomepageBKrhSXOJ05eAhBKrhSXOJ05eAhBKrhSXOJ05eAh8295680b4ce8f7de5c9990ecc009354eBKrhSXOJ05eAhNodeType\_%d0dbe915091d400bd8ee7f27f0791303%\_Medienreaktor.Site:Document.Page.HomepageBKrhSXOJ05eAhBKrhSXOJ05eAhBKrhSXOJ05eAh