Analyse des Gaz Emis - Hyphenated Techniques

PERSEUS® STA 449 F1 /F3 Jupiter® / TG 209 F1 Libra®

The Revolution in Coupling

Affordable Gas Analysis

The unrivaled instrument alliance serves as an innovative coupling technique, even for a narrow budget. It has what it takes to become an integral part of any laboratory and indispensible for its future needs.

No Liquid Nitrogen required

The DLaTGS (DLaTGS = deuterated L-alanine doped triglycine sulphate) detector used does not require any liquid nitrogen cooling. This system is therefore particularly well-suited for test runs with an autosampler (ASC) or for measurements of longer duration.

TG 209 F1 Libra® with PERSEUS® Coupling
STA 449 F1 Jupiter® with PERSEUS®® Coupling

No Separate Transfer Line

The built-in heated gas cell is directly connected to the gas outlet of the furnace via a heated tube. The low volume of this short gas path guarantees fast response and is quite advantageous in cases where condensable evolved gases are present.

Space-Saving Design

No separate FT-IR spectrometer is necessary, the footprint of the PERSEUS® coupling is identical to that of an STA 449 F1 /F3 Jupiter® / TG 209 F1 Libra® alone. This makes the coupled system ideally suited for labs in which space is scarce.

Technical Data

Heating of transfer tube
two options (temperature-controlled; using a constant voltage supply source)
Heating of gas cell
200°C in maximum,
software-controlled
Detector
DLaTGS

Gas cell length/volume:
70 mm / 5.8 ml
(no mirror inside, beam conforming design)

Spectral range of FT-IR data:
6000 cm-1 to 500 cm-1

Gas cell:
Window ZnSe, sealing Viton©

Affordable Gas Analysis

The unrivaled instrument alliance serves as an innovative coupling technique, even for a narrow budget. It has what it takes to become an integral part of any laboratory and indispensible for its future needs.

No Liquid Nitrogen required

The DLaTGS (DLaTGS = deuterated L-alanine doped triglycine sulphate) detector used does not require any liquid nitrogen cooling. This system is therefore particularly well-suited for test runs with an autosampler (ASC) or for measurements of longer duration.

TG 209 F1 Libra® with PERSEUS® Coupling
STA 449 F1 Jupiter® with PERSEUS®® Coupling

No Separate Transfer Line

The built-in heated gas cell is directly connected to the gas outlet of the furnace via a heated tube. The low volume of this short gas path guarantees fast response and is quite advantageous in cases where condensable evolved gases are present.

Space-Saving Design

No separate FT-IR spectrometer is necessary, the footprint of the PERSEUS® coupling is identical to that of an STA 449 F1 /F3 Jupiter® / TG 209 F1 Libra® alone. This makes the coupled system ideally suited for labs in which space is scarce.

Technical Data

Heating of transfer tube
two options (temperature-controlled; using a constant voltage supply source)
Heating of gas cell
200°C in maximum,
software-controlled
Detector
DLaTGS

Gas cell length/volume:
70 mm / 5.8 ml
(no mirror inside, beam conforming design)

Spectral range of FT-IR data:
6000 cm-1 to 500 cm-1

Gas cell:
Window ZnSe, sealing Viton©