Análise de Gases Desprendidos - Hyphenated Techniques

PERSEUS® STA 449 F1 /F3 Jupiter® / TG 209 F1 Libra®

The Revolution in Coupling

Affordable Gas Analysis

The unrivaled instrument alliance serves as an innovative coupling technique, even for a narrow budget. It has what it takes to become an integral part of any laboratory and indispensible for its future needs.

No Liquid Nitrogen required

The DLaTGS (DLaTGS = deuterated L-alanine doped triglycine sulphate) detector used does not require any liquid nitrogen cooling. This system is therefore particularly well-suited for test runs with an autosampler (ASC) or for measurements of longer duration.

TG 209 F1 Libra®® with PERSEUS®® Coupling
STA 449 F1 Jupiter®® with PERSEUS®®® Coupling

No Separate Transfer Line

The built-in heated gas cell is directly connected to the gas outlet of the furnace via a heated tube. The low volume of this short gas path guarantees fast response and is quite advantageous in cases where condensable evolved gases are present.

Space-Saving Design

No separate FT-IR spectrometer is necessary, the footprint of the PERSEUS®® coupling is identical to that of an STA 449 F1 /F3 Jupiter®® / TG 209 F1 Libra®® alone. This makes the coupled system ideally suited for labs in which space is scarce.

Product Information

Request information for this product.

Informação Requerida

Dados Técnicos

Heating of transfer tube
two options (temperature-controlled; using a constant voltage supply source)
Heating of gas cell
200°C in maximum,
software-controlled
Detector
DLaTGS

Gas cell length/volume:
70 mm / 5.8 ml
(no mirror inside, beam conforming design)

Spectral range of FT-IR data:
6000 cm-1 to 500 cm-1

Gas cell:
Window ZnSe, sealing Viton©