| Published: 

Thermal Expansion of Porous Metal Foams

Introduction 

In two very recent publications, measurements of the Thermal DiffusivityThermal diffusivity (a with the unit mm2/s) is a material-specific property for characterizing unsteady heat conduction. This value describes how quickly a material reacts to a change in temperature.thermal diffusivity on porous metal foams using LFA (Laser/Light Flash Analysis) were comprehensively discussed [1, 2]. The intention of this application note is to discuss another important thermophysical property of these materials: thermal expansion provided by DIL (dilatometry).

The materials investigated were open-cell foams based on AlSi7Mg (EN AC-42000) aluminum alloy, provided by Exxentis AG (Wettingen, Switzerland). The foams are created by aluminum alloy casting with crystal salt. Different pore sizes are achieved by variation of the grain size of the salt. Such foams are used as vacuum foaming molds, as thermoforming tools, for vacuum plates in vacuum tables and clamping systems, as silencers, as filters and as heat exchangers. Ultralight metal foams are also used in applications in catalysis, fuel cells, hydrogen storage and acoustic insulation [2].

Experimental

Three open-cell foams with nominal pore sizes in ranges from 0.2 to 0.35 mm (“small pores”), 0.40 to 1.00 mm (“medium pores”), and 0.63 to 4.00 mm (“large pores”) were investigated. Photos of these samples are shown as inserts in figure 1b). All foam samples had a nominal DensityThe mass density is defined as the ratio between mass and volume. density ρ = 1.09 g/cm3, or a nominal porosity of about 60%. The expansion behavior of the three porous metal foams was compared to the fully dense AlSi7Mg material with a DensityThe mass density is defined as the ratio between mass and volume. density of ρ = 2.68 g/cm3. A photo of this sample is shown as an inset in figure 1a). The DensityThe mass density is defined as the ratio between mass and volume. density of the foams was calculated as mass divided by volume. For determination of the DensityThe mass density is defined as the ratio between mass and volume. density of the fully dense sample, a DensityThe mass density is defined as the ratio between mass and volume. density balance was used. All samples were cylindrically shaped with a diameter of 12.6 mm and a thickness of 10 mm.

Measurement Conditions

The measurements were performed with a DIL 402 Expedis®Select pushrod dilatometer, equipped with a steel furnace capable of operation between -150°C and 1000°C. The system is vacuum-tight, allowing measurements to be carried out in pure inert or oxidizing atmospheres as well as under vacuum. A set of primary standards, including fused silica, sapphire, platinum, tungsten, etc., is available for length calibration. The expected expansion of the specimen and the temperature range of the measurement dictate which standard should be used. The measurements were performed with a fused silica sample holder in the temperature range from -100°C to 500°C at a heating rate of 2 K/min in a helium atmosphere. Each sample was heated twice; the results of the second heating were used to calculate the DensityThe mass density is defined as the ratio between mass and volume. density curve based on the DensityThe mass density is defined as the ratio between mass and volume. density at room temperature and the measured thermal expansion assuming isotropic expansion behavior and no mass loss during heating. In order to correct for the expansion of the sample holder and pushrod, a correction measurement with an Al2O3 reference was conducted prior to the sample measurements.

Measurement Results

Figure 1a) presents the data for the three foam samples with different pore sizes and 1b) the DensityThe mass density is defined as the ratio between mass and volume. density data of the fully dense sample. Due to the thermal expansion, the DensityThe mass density is defined as the ratio between mass and volume. density of all samples decreases with increasing temperature, showing a consistent trend. For the fully dense sample as well as for the foams, the DensityThe mass density is defined as the ratio between mass and volume. density decreases by 4.3% in the temperature range between -100°C and 500°C. Introducing porosity into a fully dense AlSi7Mg sample does not seem to significantly affect the change in DensityThe mass density is defined as the ratio between mass and volume. density with temperature. Different pore sizes in AlSi7Mg foams do not seem to have any significant effect on the DensityThe mass density is defined as the ratio between mass and volume. density behavior either.

Explore thermal analysis and rheology in polymer additive manufacturing with Dr. Natalie Rudolph, featuring a green 3D-printed sample.
1) DensityThe mass density is defined as the ratio between mass and volume. Density plotted versus temperature for a) the three AlSi7Mg foam samples with small, medium and large pores and b) the fully-dense AlSi7Mg sample
Graph comparing coefficients of thermal expansion for solid aluminum alloy and AlSi7Mg foams with varying pore sizes.
2) Coefficients of thermal expansion for the fully dense material as well as for the three AlSi7Mg foams with different pore sizes

Comparison of the Coefficient of Linear Thermal Expansion (CLTE/CTE)The coefficient of linear thermal expansion (CLTE) describes the length change of a material as a function of the temperature.CTE curves in figure 2 reveals that the curves of the fully dense sample and of the sample with large pores are, interestingly, nearly congruent. These two samples have lower overall surface areas (internal plus external) than the samples with medium and small pores and might thus exhibit more pronounced inertia versus temperature changes. Since, in dilatometry, measurements usually are performed dynamically at a specific heating rate, these samples are expected to equilibrate slower than the samples with medium and small pores and could therefore easily lag behind in their response behavior. This is a possible explanation for the slight differences in the measurement curves in figure 2, which could therefore be caused by a mixture of samplespecific and metrological effects. 

AlSiMg alloys are known to exhibit precipitation/posthardening effects, which could also play a significant role. Specific Heat Capacity (cp)Heat capacity is a material-specific physical quantity, determined by the amount of heat supplied to specimen, divided by the resulting temperature increase. The specific heat capacity is related to a unit mass of the specimen.Specific heat capacity data for the samples yielded by DSC (differential scanning calorimetry) revealed slight ExothermicA sample transition or a reaction is exothermic if heat is generated.exothermic effects in the temperature range between 250°C and 400°C [2]. The Thermal DiffusivityThermal diffusivity (a with the unit mm2/s) is a material-specific property for characterizing unsteady heat conduction. This value describes how quickly a material reacts to a change in temperature.thermal diffusivity, investigated by LFA, shows a deviation from the monotonous trend in this temperature range, too [2]. In this temperature range, the Coefficient of Linear Thermal Expansion (CLTE/CTE)The coefficient of linear thermal expansion (CLTE) describes the length change of a material as a function of the temperature.CTE curves also show extrema, probably also related to precipitation hardening. The difference in intensity of these effects could yield the  differences in the curves shown in figure 2.

Conclusion

Dilatometer measurements on a fully dense AlSi7Mg material and three AlSi7Mg foams with different pore sizes revealed similar behavior of the Coefficient of Linear Thermal Expansion (CLTE/CTE)The coefficient of linear thermal expansion (CLTE) describes the length change of a material as a function of the temperature.CTE for all samples investigated, independent of the pore size. The trend regarding the change in DensityThe mass density is defined as the ratio between mass and volume. density is about the same for all of the samples. The Thermal DiffusivityThermal diffusivity (a with the unit mm2/s) is a material-specific property for characterizing unsteady heat conduction. This value describes how quickly a material reacts to a change in temperature.thermal diffusivity of the samples, as another highly important thermophysical property, does not show such invariance to the pore size of the samples: It was found to decrease with increasing pore size.

Literature

  1. [1]
    A. Lauerer, A. Lunev, Experimental evidence of gasmediatedheat transfer in porous solids measured by theflash method, Int. J. Therm. Sci., under review, June 2022.
  2. [2]
    A. Lunev, A. Lauerer, V. Zborovskii, F. Leonard, Digitaltwin of a laser flash experiment helps to assess the thermalperformance of metal foams, Int. J. Therm. Sci., 181,107743, 2022. https://doi.org/10.1016/j.ijthermalsci.2022.107743
  3. [3]
    URL: https://slidetodoc.com/metal-foam-introduction-a-metal-foam-is-a/, visited on 06/22/2022
AI Overview
An error occurred. Please try again.