Highlights

Guarded Hot Plate - The Absolute Method for Testing Insulation Materials

Measuring devices of the GHP 721-600 mm series are robust cabinet devices which are especially suited for testing thicker specimens. Thanks to the integrated single board computer (SBC) with Windows operating system, the Lambda software and the high-resolution color touch display, the devices are very user-friendly.

Numerous interfaces such as RS232, USB and Gigabit Ethernet enable connectivity to peripheral devices and allow for fast and convenient data transfer. Connection to an external PC allows for the extensive evaluation of test results and the printing of test reports.

NETZSCH GHP 600 instrument with open doors, showcasing advanced testing technology for material analysis and research.
Thermal conductivity chart comparing materials like diamond and water, highlighting measurement methods such as Laser Flash.

Guarded Hot Plate - Principle of Operation 

The hot plate and the guard ring are sandwiched between two samples of the same material and approximately the same thickness, d. Cold plates are placed above and below the samples. All plate temperatures are controlled such that a well‐ defined, user‐selectable temperature difference, ΔT, is established between the hot and the cold plates – and thus across the sample thickness. The guard ring is maintained exactly at hot plate temperature in order to minimize lateral heat losses.

The GHP 721-600 mm is based on various relevant standards, e.g., ISO 8302, ASTM C177, EN 1946-2, EN 12664, EN 12667 and EN 12939.

Specifications

All features at a glance

GHP 721-600 mm
Measuring range0.005 to 2.0 W/(m·K), depending on material and thickness
Specimen size (L x W)

600 mm x 600 mm

dimension of the hot plate: 300 mm x 300 mm

Specimen thickness (H)
  • 1x 15 mm to 200 mm (1-specimen measurement method)
  • 2x 15 mm to 100 mm (2-specimen measurement method)
Temperature range
  • Cooling plate: -15°C to 60°C
  • Heating plate: -5°C to 70°C
Interface1x RS 232,1x Gigabit Ethernet
Dimensions (H x W x D)186 cm x 85 cm x 85 cm
Power supply110 V to 230 V, 50/60 Hz
Weight214 kg
Specimen adapter plate for 1-specimen mode, featuring a grid of 100 holes for specimen placement and electrical connection.
Specimen adapter plate for 1-specimen mode
Sensor plate with multiple contact points and connecting wire, used for precise analysis and testing in various applications.
Sensor plate

Accessories

Accessories and options for expanding upon the functions of the devices are available. The operator can adapt the different measuring tasks as needed with regard to determining the Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity using heat flow meters, guarded hot plates and guarded hot pipes. Auxiliary equipment is required for certain measurements, especially those involving materials such as insulating glass or other insulating materials, natural stone, concrete, plastics, composite materials or loose materials like granulates or flakes. In addition, suitable reference materials should be used for routinely checking and validating the devices.

Customer support representative at a computer, smiling and engaged, highlighting NETZSCH's commitment to service excellence.


Proven Excellence in Service

At NETZSCH Analyzing & Testing, we offer a comprehensive range of services globally to ensure the optimal performance and longevity of your thermoanalytical equipment. With a track record of proven excellence, our services are designed to maximize the effectiveness of your devices, extend their lifespan, and minimize downtime. 

Unlock the full potential of your equipment with our tailored solutions, backed by years of industry expertise and innovation.

Thermal testing interface displaying temperature and heat flow data for a sample in a controlled environment.

Supported Measuring Methods

  • Method with heat flow meter in accordance with ISO 8301, ASTM C518, DIN EN 1946-3, EN 12664, EN 12667 and EN 12939
  • One- or two-specimen measurement method with guarded hot plate in accordance with ISO 8302, ASTM C177, DIN EN 1946-2, EN 12664, EN 12667, EN 12939, EN 674 and ISO 10291
  • Thermal conductivity measuring device with guarded hot pipe in accordanc with DIN EN ISO 8497, DIN EN 1946-5, DIN 52613, ASTM C 534 and ASTM C 335

Functions

  • Choice of manual or automatic measuring procedure with up to 16 definable mean temperatures per measurement
  • Creation of favorites for fast access to frequently used measurement tasks
  • Display of all relevant data, measuring results, intermediate and final results as graphs and tables
  • Monitoring of relevant messages and information ∙ Safety function via error messages
  • Calculation of nominal λ value from ascertained λ90/90
  • Customized test report
  • Intuitive icons for menu functions ∙ User and administrator level
Temperature graph comparing cold (green) and warm (orange) sides over time for analysis in engineering testing.
Ferrari F1 car in pit stop, surrounded by team members in red uniforms preparing for a quick tire change.

Consultancy & Sales

Do you have further questions about the instrument or the method and would like to speak to a sales representative?

Service & Support

Do you already have an instrument and need technical support or spare parts?

Downloads and Media

AI Overview
An error occurred. Please try again.