Published on 

Shelf Life Determination of Potassium Clavulanate Via Thermogravimetry

Introduction

Potassium clavulanate (figure 1) is a salt of clavulanic acid which is a major ß-lactam antibiotic produced by the organism Streptomyces clavuligerus [1]. On its own, it is actually only capable of weak antibacterial activity against most organisms, but in combination with the antibiotic amoxicillin, it is effective against ß-lactamaseproducing staphylococcus bacteria which are resistant to amoxicillin alone [2, 3]. That is why it is an established substance in the pharmaceutical industry. 

Amoxicillin and potassium clavulanate exhibit similar Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition pathways. However, the stability of the amoxicillin-clavulanate combination mainly depends on the clavulanate, which is the more degradable of the two [4, 5]. 

The Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition of potassium clavulanate has been investigated in many papers [3, 4, 7, 12]. Generally, the substance was studied in solutions with different pH levels and in the presence of amoxicillin. It was observed that the stability of the admixture amoxicillin/clavulanic acid is affected by an increase in temperature from 25°C to 40°C [3]. On the other hand, the shelf-life of the admixture increases significantly if the pH of the solution is acidified [4]. It was also observed that in solutions, the Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition of clavulanic acid is catalyzed by the products of hydrolysis [12]. As is shown by using the HPLC method on samples stored at different temperatures and under different atmospheric conditions, the Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition of potassium clavulanate in the solid state follows another mechanism: The Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition products formed in the solid phase do not have any catalytic effect [8]. 

The Thermal StabilityA material is thermally stable if it does not decompose under the influence of temperature. One way to determine the thermal stability of a substance is to use a TGA (thermogravimetric analyzer). thermal stability can also be explored with thermogravimetry, which determines, amongst other things, the temperature at which a material starts to decompose or react [9]. The thermal decomposition of solid potassium clavulanate was characterized by means of a thermobalance coupled to an FT-IR spectrometer in [13]. In the following, thermogravimetric measurements are used to carry out kinetic studies of the Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition reaction. 

This allows for the prediction of the potassium clavulanate degradation for specific temperature and time conditions. Knowledge of the Thermal StabilityA material is thermally stable if it does not decompose under the influence of temperature. One way to determine the thermal stability of a substance is to use a TGA (thermogravimetric analyzer). thermal stability and understanding the decomposition process of potassium clavulanate in the solid state allows for optimizing its storage conditions.

Potassium clavulanate structure showcasing its molecular components and functional groups essential for antibiotic effectiveness.
1) Structure of potassium clavulanate [6]

Experimental

The TGA measurements were performed with a NETZSCH TG 209 F1 Libra® thermobalance with automatic sample changer. From the TGA-FT-IR measurement described in [13], we learned that the sample releases surface water as soon as the measurement starts. For this reason, the next measurements were carried out using closed aluminum crucibles. Just prior to the measurement, the crucible lid was automatically pierced by the piercing device of the ASC. This prevents the sample from releasing its surface water already before the actual measurement starts, which would falsify the value of the initial mass.

The sample masses were between 4.33 and 5.04 mg. The samples were heated between room temperature and 600°C at four heating rates varying from 1 K/min to 10 K/min. The measurements were carried out in a dynamic nitrogen atmosphere (40 ml/min). 

The TGA curves obtained are the basis for the kinetic evaluation of the Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition reaction. 

For this, the Kinetics Neo software (by NETZSCH-Gerätebau GmbH) was used. It allows for modelling the kinetics of single to multi-step reactions. 

This software can assign each individual step to different reaction types with kinetic parameters of their own, such as activation energy, order of reaction, and pre-exponential factor. Based on the results, Kinetics Neo is able to simulate the reaction(s) for user-specified temperature programs.

Results and Discussion

TGA Measurements 

Figure 2 depicts the TGA and DTG (first derivative) curves of the measurements on potassium clavulanate at heating rates of 1, 3, 5 and 10 K/min. The first mass-loss step, detected between room temperature and 120°C, results from the evaporation of surface water [13]. Further, the three mass-loss steps identified between 120°C and 600°C are due to the decomposition of potassium clavulanate. They are shifted to higher temperatures with increasing heating rates (kinetic influence). For example, at a heating rate of 1 K/min, the first decomposition step occurs at 167°C (DTG peak), while at a heating rate of 10 K/min, it occurs at 184°C (DTG peak). The last decomposition step becomes more pronounced with increasing heating rate: At a heating rate of 5 K/min, a DTG peak is observed at 412°C (red dashed curve), while at 10 K/min, it occurs at 417°C (black dashed curve).

Kinetic Analysis of the Thermal Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. Decomposition

The dependence of the decomposition on the heating rate allows for evaluation of the process with the help of NETZSCH Kinetics Neo software. Figure 3 shows the TGA measurement curves between 130°C and 600°C used for the kinetic evaluation. The release of surface water at temperatures lower than 130°C is not taken into consideration.

TGA and DTG analysis of potassium clavulanate in pierced crucibles at varying heating rates shows thermal stability and mass loss trends.
2) TGA measurement on potassium clavulanate in pierced crucibles at different heating rates, solid lines: TGA, dashed lines: DTG
Diagrams illustrate NBR post-vulcanization results, detailing static force, average length, relaxation modulus, and temperature changes over time.
3) Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. Decomposition of potassium clavulanate to 600°C at four different heating rates, measured TGA data
DTG curve depicts a 5-step decomposition process between 100°C and 500°C, highlighting key temperature peaks.
4) DTG curve of the measurement at 1 K/min between 100°C and 500°C. The 3 peaks and 2 shoulders indicate a 5-step decomposition process.

 

The three consecutive mass losses detected indicate at least three decomposition steps. The DTG curve of the measurement at 1 K/min presented in figure 4 shows three peaks at 167°C, 293°C and 368°C, but also two shoulders with onset temperatures at 241°C and 322°C. That is the reason why Kinetics Neo proposes a kinetic model with five consecutive steps of nth order.

The reaction rate of each step j is described by the function: Reaction Ratej = Aj · f(ej,pj) · exp[-Ej/RT]


Aj: pre-exponential factor
Ej: activation energy [J.mol-1]
T: temperature [K]
R: gas constant (8.314 J.K-1.mol-1)
f(ej,pj): function dependent on the concentration of the
initial reactant ej and the concentration of product pj

Figure 5 compares the measured TGA curves (dotted lines) with the calculated curves (solid lines) of the chosen 5-step model. A high correlation coefficient, of >0.999, is achieved between the measured and calculated data. 

Table 1 summarizes the results of the kinetic evaluation for each step. The theoretical mass loss is calculated by multiplying the contribution of the reaction step to the decomposition with the total mass loss occurring during decomposition. 

The first decomposition step, A→B, is associated with a calculated mass loss of 11.9% and corresponds to the experimental values of 11%. The mass loss for the last step, E→F, amounts to 13.9%. This is slightly higher than the experimental value of 11 - 12%. It means that the last mass-loss step starts earlier (< 360°C). The total mass loss of steps B→C, C→D and D→E is 36.9% and corresponds to the complex decomposition process around 300°C (DTG peak) in figure 2.

Kinetic evaluation graph of potassium clavulanate decomposition, showing measured vs. calculated curves across different reaction rates.
5) Kinetic evaluation of the decomposition of potassium clavulanate. Dotted lines: measured curves; solid lines: calculated curves based on a five-step reaction of nth order. The correlation coefficient between measured and calculated curves amounts to >0.999.

 

Tab 1: Kinetic parameters of the thermal degradation of potassium clavulanate

Reaction step

A → B

B → C

C → D

D → E

E → F

Activation energy [kJ/mol]

265.1

240.8

260.5

179.8

166.5

Pre-exponential factor

28.6

21.6

21.7

13.3

10.5

Reaction order

3.6

2.1

1.8

1.6

3.4

Contribution

0.190

0.099

0.244

0.246

0.222

Theoretical mass loss

11.9%

6.2%

15.3%

15.4%

13.9%

The good correlation of the measurements with nth order reactions confirms the conclusions drawn in [8] that contrary to its decomposition behavior in solutions, the decomposition of potassium clavulanate in a solid state is not auto-catalyzed. 

The kinetic evaluation was carried out with a high correlation coefficient and thus a high level of agreement between the measured and simulated TGA curves, so that predictions about the long-term behavior under different storage temperatures are possible. As an example, figure 6 shows the mass change versus time based on the 5-step model with consecutive steps; it represents the prediction of the decomposition of potassium clavulanate for various temperatures between 80°C and 150°C in a nitrogen atmosphere. With increasing temperature, the decomposition increases. This effect can already be observed at a storage temperature of 90°C (green curve at the upper end of the graph – figure 6). 

Figure 7 depicts the stability in an inert atmosphere of the drug over the course of 5 years, for temperatures between 20°C and 80°C. It seems that no significant mass loss occurs in the prediction for temperatures up to 60°C.

Graph depicting the decomposition predictions of potassium clavulanate over 24 months at varying temperatures in a nitrogen atmosphere.
6) Predictions of the decomposition of potassium clavulanate in a nitrogen atmosphere for different temperatures over a time period of 2 years
Five-year forecast graph showing potassium clavulanate mass decomposition at temperatures between 20°C and 80°C in nitrogen atmosphere.
7) Five-year forecast of the decomposition process of potassium clavulanate in a nitrogen atmosphere between 20°C and 80°C
Comparison graph showing measured vs. predicted mass change of potassium clavulanate during heating to 200°C over time.
8) Comparison of the measured and predicted mass change of potassium clavulanate during heating to 200°C and IsothermalTests at controlled and constant temperature are called isothermal.isothermal segment; the release of surface water is not monitored.

 

A reminder should be inserted here that the decomposition kinetic was carried out on a dry sample. However, water has great influence on the decomposition of potassium clavulanate: Storage in a humid atmosphere shifts its decomposition to lower temperatures [10]. J. Cieleka-Piontek shows that potassium clavulanate samples decompose faster if exposed to increased air humidity than when they are exposed to dry air and suggests that the attack of a water molecule on the carbonyl group of the ß-lactam ring induces thermolysis [8]. 

In order to validate the kinetic model calculated by Kinetics Neo for prediction of the decomposition behavior under IsothermalTests at controlled and constant temperature are called isothermal.isothermal conditions, a potassium clavulanate sample of 9.23 mg was heated to 200°C and then kept IsothermalTests at controlled and constant temperature are called isothermal.isothermal for two hours. Monitoring of the measurement started at 120°C in order to exclude the mass-loss effect of the release of surface water. 

Figure 8 compares the mass losses determined via measurement to those determined via prediction (Kinetics Neo). The comparison shows the good agreement between the two curves and thus the reliability of the calculation.

Conclusion

The kinetics of the thermal decomposition of potassium clavulanate in a solid state under nitrogen were investigated by means of thermogravimetry and Kinetics Neo. A high level of correlation between measured and simulated data can be obtained by using a consecutive five-step kinetic model where each step is of nth order. This allows for predictions of the storage behavior under different temperatures, temperature profiles and periods. 

The results are validated by comparing the TGA measurement under a specified temperature profile including IsothermalTests at controlled and constant temperature are called isothermal.isothermal segment to the predictions calculated by Kinetics Neo.

Literature

  1. [1]
    Clavulanic acid: A review, Biotechnology Advances, Parag S. Saudagar, Shrikant A. Survase, Rekha S. Singhal Volume 26, Issue 4, July–August 2008, Pages 335-351
  2. [2]
    PubChem Open Chemistry Database, Compound summary for CID 23665591
  3. [3]
    Stability study of 6 brands of amoxicillin trihydrate and clavulanic acid oral suspension present in Yemen markets, Mahmoud Mahyoob Alburyhi, Abdulwali Ahmad Siaf and Maged Alwan Noman, Journal of Chemical and Pharmaceutical Research, 2013, 5(5):293-296
  4. [4]
    Improving the stability of potassium clavulanate in admixture with amoxicillin, Elena M. Vega, Ruben H. Manzo and Nancy Sola, Hospital Pharmacist, May 2008, Vol. 15, pages 183-185
  5. [5]
    The stability of amoxicillin trihydrate and potassium clavulanate combination in aqueous solutions, Ákos Jerzsele and Gábor Nagy, Acta Veterinaria Hungarica 57 (4), pp. 485–493 (2009)
  6. [6]
    https://commons.wikimedia.org/wiki/File:Potassium_clavulanate_structure.svg
  7. [7]
    Kinetic studies on clavulanic acid degradation, Patricia A. Bersanetti, Renata M.R.G. Almeida, Marlei Barbozac, Maria Lucia G.C. Araujo, Carlos O. Hokka, Biochemical Engineering Journal, Volume 23, Issue 1, March 2005, Pages 31-36
  8. [8]
    Solid-state stability and compatibility studies of potassium clavulanate, Judyta Cieleka-Piontek, Magdalena Paczkowska, Przemyslaw Zalewski, Kornelia Lewandowska and Boleslaw Barszcz, Pharmaceutical Development and Technology, November 2013
  9. [9]
  10. [10]
    NETZSCH Application Note AN 117, Influence of Humidity on the Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. Decomposition of Potassium Clavulanate, https://www.NETZSCH.com/an117_en
  11. [11]
    Spektroskopische Methoden in der organischen Chemie, Manfred Hesse, Herbert Meier, Bernd Zeh, Georg Thieme Verlag Stuttgart New York
  12. [12]
    Clavulanic acid decomposition is catalyzed by the compound itself and by its decomposition products, Simone Brethauer, Martin Held, Sven Panke, Journal of Pharmaceutical Sciences, August 2008, Volume 97, Issue 8, Pages 3451-3455
  13. [13]
    NETZSCH Application Note AN 118, TGA-FT-IR for Better Understanding the Degradation of Potassium Clavulanate, https://www.NETZSCH.com/an118_en
AI Overview
An error occurred. Please try again.