NETZSCH TCT 716 Lambda thermal analysis instrument with two measurement chambers, featuring digital controls and sleek design.

Highlights

TCT 716 Lambda - Between Classical Heat Flow Meters and Laser Flash Analyzers

The TCT 716 Lambda offers the capability of analyzing specimens with optimum dimensions: smaller than conventional HFM and larger than LFA. This enables investigations on homogeneous and inhomogeneous materials with Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity values ranging from low to medium, for example polymers, composites, glass, ceramics, some metals, etc.

The robust design of the TCT 716 LambdaThermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity tester offers easy, uncomplicated handling of the software and hardware. The guarded heat flow meter (GHFM) is fully software-controlled, including mean temperature and applied force. The software also allows for an unlimited number of steps in test cycles for best performance

This GHFM has a left and right test stack; which enables tests on a single specimen or simultaneous tests on two specimens. Each stack is independent of the other in terms of clamping force and specimen thickness. Both stacks can be operated across the entire temperature range from -10°C to 300°C. This arrangement not only increases sample throughput, but also allows for more data to be collected in less time.

The system provides precise temperature control with a resolution of 0.1°C. It is equipped with multiple high-resolution detectors (RTD), which allow for accurate measurement of the thermal gradient across stack and specimen thickness.

NETZSCH TCT 716 Lambda thermogravimetric analyzer with digital interface, showcasing precision instruments for thermal analysis.

Cost-Effective Cooling

CO2 is a natural refrigerant that provides sustainable and energy-efficient cooling in everything from warehouses to ice machines – including the TCT 716 Lambda! 

CO2 features unique thermophysical properties:

  • Very good heat transfer coefficient
  • High energy content
  • Relatively insensitive to pressure losses
  • Very low viscosity Unlike other GHFMs, this design allows for CO2 cooling for optimal temperature control.

There is no longer any need for an expensive chiller unit. In addition, forced cooling of the instrument is possible and CO2 consumption above ambient temperature is low.

Please accept Marketing Cookies to see that Video.

Thermal conductivity is a measure of a material's ability to conduct heat. It quantifies how well heat can move through a substance. The most common method for measuring Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity is the steady-state method, also known as the heat flow meter method.

In this method, a sample of the material with known dimensions is placed between two plates of different temperatures. One plate is heated, while the other is cooled, creating a temperature gradient across the material. Heat flows through the sample from the hot plate to the cold plate. The rate of heat transfer (heat flux) and the temperature difference across the sample are measured.

Using Fourier's law of heat conduction, which relates the heat flux, temperature gradient, and Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity of the material, the Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity of the sample can be calculated. This calculation accounts for factors such as the dimensions of the sample and the thermal resistance at the interface between the sample and the plates.

By repeating the measurements with different samples and under various conditions, the Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity of the material can be accurately determined. This information is crucial for evaluating the insulation properties of materials used in building construction, electronics, and various other applications where heat transfer is a concern.

Graph illustrating thermal conductivity ranges of various materials, with color-coded sections from vacuum insulation to diamond.
Schematic diagram of TCT 716 Lambda, showcasing dual specimen measurement with heat flow direction indicators.
Schematic diagram of TCT 716 Lambda with measurement capability of two specimens

TCT 716 Lambda - Principle of Operation 

The operator measures the thickness of the test specimen(s), and places them between two heated plates controlled at different temperatures. Temperature sensors (RTD) are mounted just below the plate surfaces for measuring the temperature drop across the specimen. Similar sensors are also embedded in the upper and lower stacks (metering area 51 mm) to measure the heat flow through the specimen. Once steady-state condition has been achieved, these signals are collected for calculating Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity. The software indicates thermal equilibrium. After the indication of thermal equilibrium, the measurement is carried out.

A laboratory technician in a white lab coat operates a modern analytical device, focusing on precision measurements.

NEW: Contract Testing Service with TCT 716 Lambda 

The TCT 716 Lambda offers the capability of analyzing specimens with optimum dimensions: smaller than conventional HFM and larger than LFA. This allows testing of homogeneous and inhomogeneous materials with low to medium Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity, such as polymers, composites, glass, ceramics, some metals, etc.

Discuss with our experts the best possible method for your sample. Test now and take advantage of our special contract testing offer!

NETZSCH offers more products that support you in measuring Thermal Conductivity:

  • TDW 4240
    TDW 4240

    Hotbox Test Chamber for testing construction materials (windows, profiles, doors, domes, brick walls etc.)

    • Measuring Range: R: 0.10 to 8.00 m²·K/W, U: 0.12 to 3.70 W/(m²·K)
    • Specimen thickness (H): up to 560 mm
  • TLR 1000
    TLR 1000

    Measuring Device with guarded hot pipe for pipe insulations 

    • Temperature range: test chamber: -15°C to 140°C, hot pipe: 20°C to 200°C
    • Measuring range: 0.001 W/(m·K) up to 0.25 W/(m·K)
  • LFA 717 HyperFlash
    LFA 717 HyperFlash®

    A fast, non-contact method for determining thermal diffusivity

    • Temperature range: -100°C to 500°C
    • Simultaneous measurement of up to 16 samples
    • Broadest sample holder and sample material range

Specifications

 TCT 716 Lambda
General
StandardsBased on ASTM E1530
OperationExternal PC, minimum i5 or equivalent, 500 GB, 2x USB 3.0 (not included)
Automated instrument calibrationYes; reference materials: fused silica; pyroceram and stainless steel
Testing chamberMotorized door opening/closing mechanism, interlocked
Measurement data
Thermal resistance range0.001 ... 0.030 m2·K/W
Thermal conductivity range0.1 … approx. 30 W/(m·K) (using proper sample thicknesses)
Thermal conductivity accuracy±3% to 5% deviation* from literature value (depending on the accuracy of calibration material)
Thermal conductivity repeatability±2% (precision; measurement of the same sample in the same device after sample out/in between measurements
Measurement times for different material typesIn general, t < 2 hours/point, depending on range, number of temperature steps and conductivity
Number of set pointsFree-selectable number of programmable test temperatures; typically full range test includes 5 to 6 test temperatures max.
Number and type of temperature sensorsPremium RTD class A, in protective capsule, 14 total/instrument, resolution: 0.01°C
Metering area of the plates51 mm, round, full cross section
Sample Dimensions
Sample shapesRound
Sample dimensionsø 50,8 mm (2 in); height up to 31.8 mm (1¼ in)
Sample conditionSolid
Number of samplesUp to 2; independent of type, identical thermal cycles
Contact force and load control
Variable contact forceProgrammable for incompressible materials
Contact pressure/accuracy35, 70, 175, 350 kPa
Load controlAutomatic
Temperature
Temperature
  • Max. hot plate temperature: 350°C
  • Sample mean temperature range: -10°C to 300°C
Temperature gradientTypically 30 K, variable
Cooling systemLiquid CO2
RTD resolution±0.05%, class A RTD, approx. ±0.01°C resolution
Locations of temperature measurementSpecific locations along stack, consisting of upper plate/sample/lower plate, heat sink
Instrument Dimensions
Dimensions and weight460 mm (18”) wide, 630 mm (25”) deep, 510 mm (20”) high, 80 lbs (without CO2 cylinder)
CO2 cylindermandatory for operation (not included) 

*Depending on the accuracy of calibration material and specimen properties

A collection of circular objects in various colors and finishes, showcasing sleek design and materials for product analysis.
Three textured circular stones in gray, pink, and brown, ideal for home decor or aesthetic projects.

Samples

Typically, the TCT 716 Lambda allows for measurements on round solid specimens in the low- and medium-conductive range such as polymers (filled and unfilled) and low-conductive ceramics and metals including porous specimens. An important feature of the instrument is that there are no temperature sensors embedded into the sample. Sample preparation is according to ASTM E1530. For solid specimens, a thermal interface paste is used to improve the thermal contact with the instrument plates

Advantages of the GHFM

The GHFM provides a reliable and precise method for measuring the thermal conductivity and thermal resistance in a wide variety of solids, thus contributing to materials science research and product development.

  • High accuracy: uncertainties typically < 3%
  • Non-destructive test: The materials to be tested can be measured as made without destroying or otherwise altering them
  • Wide range of materials: metals, polymers, ceramics, composites, etc.
  • Specimen dimensions: 50.8 mm (2inch) in diameter, up to 31.8 mm thickness – advantageous for inhomogeneous samples
  • Easy to use: typically only  minimal training required
NETZSCH TCT 716 Lambda thermal conductivity tester with illuminated specimen chambers, designed for precise analysis in material testing.
Customer support representative at a computer, smiling and engaged, highlighting NETZSCH's commitment to service excellence.


Proven Excellence in Service

At NETZSCH Analyzing & Testing, we offer a comprehensive range of services globally to ensure the optimal performance and longevity of your thermoanalytical equipment. With a track record of proven excellence, our services are designed to maximize the effectiveness of your devices, extend their lifespan, and minimize downtime. 

Unlock the full potential of your equipment with our tailored solutions, backed by years of industry expertise and innovation.

Related Devices

  • GHP 721-600 mm
    GHP 721-600 mm

    Guarded Hot Plate with touch display – for specimen dimensions up to 600 mm x 600 mm 

    • Measuring range: 0.005 to 2.0 W/(m·K), depending on material and thickness
    • Specimen size (L x W): 600 mm x 600 mm
    • dimension of the hot plate: 300 mm x 300 mm
  • GHP 721S-900 mm
    GHP 721S-900 mm

    Guarded Hot Plate with a Tiltable Test Chamber

    • Measuring range: 0.005 to 2.0 W/(m·K), depending on material and thickness
    • Specimen size (L x W): 900 mm x 900 mm variable, according to the dimension of the hot plate: 200 mm x 200 mm up to 500 mm x 500 mm
  • LFA 467 HyperFlash
    LFA 467 HyperFlash®

    New Dimensions in the Measurement of Thermal Diffusivity and Conductivity

    • Contactless measurement of the temperature increase with IR detector
    • integrated automatic sample changer for up to 16 samples
    • Measurements from -100°C to 500°C with one furnace

     

Formula 1 pit crew servicing a red racing car, showcasing teamwork and speed during a high-pressure race scenario.

Consultancy & Sales

Do you have further questions about the instrument or the method and would like to speak to a sales representative?

Service & Support

Do you already have an instrument and need technical support or spare parts?

Downloads and Media

Videos

Please accept Marketing Cookies to see that Video.

More Videos and Webinars about the NETZSCH TCT 716 Lambda

Please accept Marketing Cookies to see that Video.

Get comprehensive insight into the technical specifications, the application fields and the unique advantages the TCT 716 Lambda!

Please accept Marketing Cookies to see that Video.

How to Start a Measurement? This is a step-by-step instruction of a measurement definition incl. preparing and inserting the specimens, software handling
AI Overview
An error occurred. Please try again.