Oxidative Stability of Polymers: The OIT Test

Introduction

Degradation of a polymer goes hand in hand with worsening of its properties and shortening its lifetime, so that stabilizers are usually added to avoid oxidative Réaction de DécompositionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition or to slow it down. An easy and quick method of measuring the stability of polymers against oxidative ContrainteLa Contrainte est définie par un niveau de force appliquée sur un échantillon d’une section bien définie. (Contrainte = force/surface). Les échantillons qui possèdent une section rectangulaire ou circulaire peuvent être comprimés ou étirés. Les matériaux élastiques comme les élastomères peuvent être étirés jusqu’à 5 à 10 fois leur longueur initiale.stress is to perform OxidationOxidation can describe different processes in the context of thermal analysis.oxidation induction tests, also called Oxidative-Induction Time (OIT) and Oxidative-Onset Temperature (OOT)Oxidative Induction Time (isothermal OIT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition. Oxidative-Induction Temperature (dynamic OIT) or Oxidative-Onset Temperature (OOT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition.OIT, with a differential scanning calorimeter (DSC). These tests are internationally recognized and described in different standards, e.g., ASTM D3895-92, ASTM D6186, EN 728 and ISO 11357-6.

Measurement Conditions

In the following, Oxidative-Induction Time (OIT) and Oxidative-Onset Temperature (OOT)Oxidative Induction Time (isothermal OIT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition. Oxidative-Induction Temperature (dynamic OIT) or Oxidative-Onset Temperature (OOT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition.OIT tests are carried out on two different polyethylene samples using the DSC 300 Caliris® Classic. In order to ensure good repeatability of the DSC curves, the samples were prepared exactly the same way and the measurements performed with crucibles without lid. In fact, this type of test is very sensitive to sample preparation and measurement parameters, and even the hole size in the lid would influence the results. 

The test consists of three different steps: 

Table 1 summarizes the measurement conditions.

Table 1: Conditions of the Oxidative-Induction Time (OIT) and Oxidative-Onset Temperature (OOT)Oxidative Induction Time (isothermal OIT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition. Oxidative-Induction Temperature (dynamic OIT) or Oxidative-Onset Temperature (OOT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition.OIT test

DeviceDSC 300 Caliris® Classic
CrucibleConcavus® (aluminum), open
Sample mass10.2 ± 0.2 mg
Temperature program

25… 220°C, 20 K/min, nitrogen (40 ml/min)

220°C, 3 min, nitrogen (40 ml/min)

220°C, synthetic air (50 ml/min)

Test Results

Figure 1 depicts the test results. The endothermal peak detected during heating is due to melting of the polyethylene. Oxidative degradation is characterized by an ExothermiqueA sample transition or a reaction is exothermic if heat is generated.exothermal effect during the isothermal segment. Here, it takes place 35.9 to 36.1 min (tangential evaluation) after switching the gas to synthetic air for sample 1 (red and pink curves). Sample 2 exhibits a better oxidative stability with an Oxidative-Induction Time (OIT) and Oxidative-Onset Temperature (OOT)Oxidative Induction Time (isothermal OIT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition. Oxidative-Induction Temperature (dynamic OIT) or Oxidative-Onset Temperature (OOT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition.OIT value of more than 10 min higher (blue and dark blue curves). The curves show the excellent repeatability of the measurements.

1) Oxidative-Induction Time (OIT) and Oxidative-Onset Temperature (OOT)Oxidative Induction Time (isothermal OIT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition. Oxidative-Induction Temperature (dynamic OIT) or Oxidative-Onset Temperature (OOT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition.OIT measurements on 2 polyethylene samples

Conclusion

An OIT test allows for easy and fast characterization of the oxidative stability of polymers and comparison of their thermo-oxidative performance. Careful sample preparation combined with the outstanding measurement performance of the DSC 300 Caliris® Classic and the automatic evaluation feature of Proteus® ensure good repeatability of the results.