Published on 

Oxidative Stability of Polymers: The OIT Test

Measurement Conditions

In the following, Oxidative-Induction Time (OIT) and Oxidative-Onset Temperature (OOT)Oxidative Induction Time (isothermal OIT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition. Oxidative-Induction Temperature (dynamic OIT) or Oxidative-Onset Temperature (OOT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition.OIT tests are carried out on two different polyethylene samples using the DSC 300 Caliris®Classic. In order to ensure good repeatability of the DSC curves, the samples were prepared exactly the same way and the measurements performed with crucibles without lid. In fact, this type of test is very sensitive to sample preparation and measurement parameters, and even the hole size in the lid would influence the results. 

The test consists of three different steps: 

Table 1 summarizes the measurement conditions.

Table 1: Conditions of the Oxidative-Induction Time (OIT) and Oxidative-Onset Temperature (OOT)Oxidative Induction Time (isothermal OIT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition. Oxidative-Induction Temperature (dynamic OIT) or Oxidative-Onset Temperature (OOT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition.OIT test

DeviceDSC 300 Caliris®Classic
CrucibleConcavus® (aluminum), open
Sample mass10.2 ± 0.2 mg
Temperature program

25… 220°C, 20 K/min, nitrogen (40 ml/min)

220°C, 3 min, nitrogen (40 ml/min)

220°C, synthetic air (50 ml/min)

Test Results

Figure 1 depicts the test results. The endothermal peak detected during heating is due to melting of the polyethylene. Oxidative degradation is characterized by an ExothermicA sample transition or a reaction is exothermic if heat is generated.exothermal effect during the IsothermalTests at controlled and constant temperature are called isothermal.isothermal segment. Here, it takes place 35.9 to 36.1 min (tangential evaluation) after switching the gas to synthetic air for sample 1 (red and pink curves). Sample 2 exhibits a better oxidative stability with an Oxidative-Induction Time (OIT) and Oxidative-Onset Temperature (OOT)Oxidative Induction Time (isothermal OIT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition. Oxidative-Induction Temperature (dynamic OIT) or Oxidative-Onset Temperature (OOT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition.OIT value of more than 10 min higher (blue and dark blue curves). The curves show the excellent repeatability of the measurements.

OIT measurements displayed for two polyethylene samples, with time, temperature, and flow rates noted on the graph.
1) Oxidative-Induction Time (OIT) and Oxidative-Onset Temperature (OOT)Oxidative Induction Time (isothermal OIT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition. Oxidative-Induction Temperature (dynamic OIT) or Oxidative-Onset Temperature (OOT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition.OIT measurements on 2 polyethylene samples

Conclusion

An Oxidative-Induction Time (OIT) and Oxidative-Onset Temperature (OOT)Oxidative Induction Time (isothermal OIT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition. Oxidative-Induction Temperature (dynamic OIT) or Oxidative-Onset Temperature (OOT) is a relative measure of the resistance of a (stabilized) material to oxidative decomposition.OIT test allows for easy and fast characterization of the oxidative stability of polymers and comparison of their thermo-oxidative performance. Careful sample preparation combined with the outstanding measurement performance of the DSC 300 Caliris®Classic and the automatic evaluation feature of Proteus® ensure good repeatability of the results.

AI Overview
An error occurred. Please try again.