Highlights

Dedicated to all-sized specimens equipped with great features.

Our Heat Flow Meter HFM 446 Lambda Medium combines innovative features:

Our SmartMode streamlines measurement, evaluation, and reporting processes, empowering operators with intuitive tools such as AutoCalibration, wizards, user-defined methods, and detailed reports. Equipped with dual heat flux transducers, our instrument ensures precision and sensitivity in monitoring heat flow to and from specimens. Calibration with reference materials of known Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity enhances accuracy, while various calibration options further boost precision.

In addition to measuring Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity, our hardware and software enable the determination of Specific Heat Capacity (cp)Heat capacity is a material-specific physical quantity, determined by the amount of heat supplied to specimen, divided by the resulting temperature increase. The specific heat capacity is related to a unit mass of the specimen.specific heat capacity (Specific Heat Capacity (cp)Heat capacity is a material-specific physical quantity, determined by the amount of heat supplied to specimen, divided by the resulting temperature increase. The specific heat capacity is related to a unit mass of the specimen.cp), providing a comprehensive analysis of thermal properties. Furthermore, the instrument prioritizes resource conservation with Eco-Mode, allowing for energy-saving standby and quick measurement startup in Idle-Mode. Users can easily customize activation timings using the scheduler, promoting efficiency in operations.

Thin film ribbon cable thermal conductivity testing

Saving and Efficient use of Energy

Today, global attention towards saving and efficiently using energy has never been higher. Industries and academia worldwide are actively researching ways to conserve energy and utilize alternative resources. Among the key focuses are insulation materials and thermal efficiency in buildings, holding vast potential. Ensuring high-quality manufacturing and strict performance control of these materials is paramount.

Various standards and guidelines govern these products to guarantee their efficacy, given the massive production volumes globally. Our latest offering, the HFM 446 Lambda Eco-Line, ensures peak energy efficiency in measuring Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.Thermal Conductivity.

Thermal conductivity is a measure of a material's ability to conduct heat. It quantifies how well heat can move through a substance. The most common method for measuring Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity is the steady-state method, also known as the heat flow meter method.

In this method, a sample of the material with known dimensions is placed between two plates of different temperatures. One plate is heated, while the other is cooled, creating a temperature gradient across the material. Heat flows through the sample from the hot plate to the cold plate. The rate of heat transfer (heat flux) and the temperature difference across the sample are measured.

Using Fourier's law of heat conduction, which relates the heat flux, temperature gradient, and Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity of the material, the Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity of the sample can be calculated. This calculation accounts for factors such as the dimensions of the sample and the thermal resistance at the interface between the sample and the plates.

By repeating the measurements with different samples and under various conditions, the thermal conductivity of the material can be accurately determined. This information is crucial for evaluating the insulation properties of materials used in building construction, electronics, and various other applications where heat transfer is a concern.

Thermal conductivity comparison chart showing various materials from vacuum insulation to diamond, indicating their conductivity values.
Schematic of a thermal conductivity testing setup, detailing components like heat sinks, Peltier systems, and test sample direction.

The HFM is an exact, fast and easy-to-use instrument for measuring the low thermal conductivity λ of insulation materials.

In a heat flow meter (HFM), the test specimen is placed between two heated plates controlled to a user-defined mean sample temperature and temperature gradient to measure heat flowing through the specimen. The sample thickness L is measured by an internal thickness gauge. Alternatively, the user can enter and drive to the desired thickness, which is of particular interest for compressible samples. The heat flow Q through the sample is measured by two calibrated heat flux transducers covering a large area of both sides of the specimen.

After reaching thermal equilibrium, the test is done. The heat flux transducer output is calibrated using a reference standard. For the calculation of the thermal conductivity λ and the thermal resistance R, the average heat flux Q/A, the sample thickness L, and the temperature gradient ΔT are used, in accordance with Fourier’s Law (see formulas on the right). The thermal transmittance, also known as U-value, is the reciprocal of the total thermal resistance. The lower the U-value, the better the insulating ability.

Material Parameter Thermal Conductivity

  • What is the heating/cooling load of a building?
  • How does this change with the weather and how can you improve it?
  • How can you optimize the heat transfer from an electronic component?
  • How do you design a heat exchanger system to achieve the required efficiency and what are the best materials to use?

To answer questions like these, material properties such as Thermal DiffusivityThermal diffusivity (a with the unit mm2/s) is a material-specific property for characterizing unsteady heat conduction. This value describes how quickly a material reacts to a change in temperature.thermal diffusivity and thermal conductivity must be known. NETZSCH offers various thermal conductivity testing instruments covering nearly all possible applications and temperature ranges.

For the analysis of lower-conductivity materials such as fiber insulations or vacuum insulation panels, NETZSCH stands out with various types of heat flow meters (HFM) for diverse sample dimensions and temperature ranges.

Cross-section of a thermal testing model showing hot and cold plates, Peltier elements, and heat flux transducers for temperature control.
Cross-section model
Thermocouples mounted on a sleek black and gold testing device against a vibrant teal background, ideal for temperature measurement.
Thermocouples
Dr. Ragy Ragheb, a technical specialist, smiles confidently in a professional setting, highlighting expertise in material characterization.
Thermocouples are placed in the specimen's center

The most important role here is played by the material parameter thermal conductivity (amount of heat per second flowing through a material layer of a thickness of 1 meter and an area of 1 m² when the temperature difference amounts to 1 K). The thicker the material layer through which the heat flows, the higher the thermal resistance (R-Value) that the material layer presents to the quantity of heat to be transported. The reciprocal value of the thermal resistance is the thermal transmittance (U-Value), usually specified for structural components.

No matter whether for expanded polystyrene (EPS), extruded polystyrene (XPS), PU rigid foam, mineral wool, bloated perlite or foam glass, cork, fleece or natural fiber materials – no matter whether for building materials containing phase change materials, aerogels, concrete, plaster or polymers or even high performing insulation materials like vacuum insulation panels (VIP's) – the new HFM 446 Lambda Eco-Line features a new standardized method for the measurement of thermal conductivity which is equally applicable in research and development and in quality assurance.

A temperature gradient is set between two plates through the material to be measured. By means of two highly accurate heat-flow sensors in the plates, the heat flow into the material and out of the material, respectively, is measured. If the state of equilibrium of the system is reached and the heat flow is constant, the thermal conductivity can be calculated with the help of the Fourier equation as long as the measurement area and thickness of the sample are known.

The HFM 446 Lambda Medium excels in Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.Thermal Conductivity Measurements across a range of materials including insulating materials, polymers, phase-change materials, aerogels, and non-wovens.

With two flexible measurement options, users can connect with a computer via our SmartMode software or opt for standalone use with an integrated printer. Setup is quick and easy, thanks to factory calibration with certified reference materials (NIST SRM 1450d).

The instrument offers fast measurements, up to 40% quicker than previous versions, with the added benefit of an Eco-Mode for reduced energy consumption. Closed test chambers ensure optimal test conditions, minimizing environmental influences and condensation risks.

Innovative features include sample thickness and parallelism measurement using a two-axis inclinometer, high throughput facilitated by motorized plate and door movement, and the ability to extend the thermal conductivity range using external thermocouples.

Advanced functionalities like the "Drive-to-thickness" feature and Heat Flux Transducer (HFT) MultiCalibration enhance accuracy and precision. Complete QA documentation and compliance to standards are ensured with Stability Configuration Management.

Improved instrument handling and compatibility with multiple operating systems and languages make our device user-friendly and accessible to all. Additionally, Specific Heat Capacity (cp)Heat capacity is a material-specific physical quantity, determined by the amount of heat supplied to specimen, divided by the resulting temperature increase. The specific heat capacity is related to a unit mass of the specimen.specific heat capacity (Specific Heat Capacity (cp)Heat capacity is a material-specific physical quantity, determined by the amount of heat supplied to specimen, divided by the resulting temperature increase. The specific heat capacity is related to a unit mass of the specimen.cp) measurement is supported based on the step method, offering comprehensive thermal analysis capabilities.

The HFM 446 Lambda Series is based on various relevant standards, e.g., ASTM C518, ISO 8301, JIS A1412, DIN EN 12664* and DIN EN 12667.

* not for HFM 446 Lambda Large

Specifications

HFM 446 Lambda Medium
StandardsASTM C518, ISO 8301, JIS A1412, DIN EN 12667, DIN EN 12664
TypeStand-alone, with integrated printer
Thermal conductivity range

0.002 to 2 W/(m·K)**

Small and Medium: 2.0 W/(m·K) achievable with optional instrumentation kit, recommended for hard materials and those with higher thermal conductivity

Performance data:

  • Accuracy: ± 1% to 2%
  • Repeatability: ± 0.25 %
  • Reproducibility: ± 0.5%

→ All performance data is verified with NIST SRM 1450 D (thickness 25 mm)

Plate temperature range-20°C to 90°C, optional for the HFM 446 Lambda Medium: -30° to 90°C
Air-tight systemSample compartment with possibility to introduce purge gas
Metering area heat flux transducer102 mm x 102 mm
Chiller systemExternal; constant temperature setpoint over plate temperature range
Plate temperature controlPeltier system
Plate motionMotorized
Plate thermocouplesThree thermocouples on each plate, type K (two extra thermocouples with instrumentation kit)
Thermocouple resolution± 0.01°C
Number of setpointsUp to 99
Specimen sizes (max.)305 mm x 305 mm x 105 mm
Variable load/ contact force

0 to 1930 N (21 kPa on 305 x 305 mm²)

Force-controlled adjustment of the contact force or the desired thickness, and thus DensityThe mass density is defined as the ratio between mass and volume. density, of compressible materials

Thickness determination
  • Automatic measurement of mean sample thickness
  • Four-corner thickness determination via inclinometer
  • Compliance to non-parallel specimen surfaces
Software features

** Please note: In the very low thermal conductivity range, accuracy of Lambda (λ) values can be restricted

Accessories and more:

The HFM system is delivered calibrated with an NIST-certified or IRRM reference standard of known thermal conductivity. This establishes precise correlation between the signal output of the transducers and the actual heat flow. Thermal resistance and thermal conductivity are determined once the userdefined stability criteria are met. Of course, the customer can also use his own reference materials. For the above-mentioned reference specimens, an accuracy of ± 1% can be easily achieved.

A close-up of a shelf inside a green acoustic panel, featuring a yellow foam layer for sound absorption.
Inserting the reference material; various reference materials can be used

The design of the testing chamber of the HFM 446 Lambda Series minimizes influences from the environment and reduces condensation effects inside the testing chamber and on the plate surfaces. Optionally, the testing chamber can be purged with dry gas.

Software

All software highlights at a glance

Thermal conductivity test results displayed on NETZSCH software and printed report, showcasing measurement graphs and data.

Highest Usability

SmartMode is the user-friendly, smoothly running user interface of the HFM Proteus® software. It is characterized by a logical structure which quickly gives a clear overview of the current measurement status and provides various report and export possibilities. After completing the test, all relevant results can be directly printed out by the integrated printer or a report can be created by the software when a PC is connected.

Calibration in Next to no Time

For calibration purposes, the thermal conductivity values of the most common certified reference materials, such as NIST SRM 1450d, are already stored in the software. However, AutoCalibration also offers the ability to create calibration curves for any user-defined material on the basis of up to 99 freely selectable temperatures.

Graph depicting apparent shear rate (γ̇app) versus inverse capillary radius (1/R) demonstrating polymer behavior under varying conditions.

Find out even more:

The “Wizards” button allows for manual parameter input, while the User Methods button retrieves parameter sets defined by the user beforehand. Such “User Methods” can also be transferred to “Favorites” for faster access if they are used frequently.

The button “Reports” allows for reports to be generated quickly and easily by granting access to various templates; one of these templates meets all of the requirements stipulated by ASTM C518. Each report can be adapted to the customer‘s own corporate identity. In addition, data can easily be exported into either Word or Excel format with just a few mouse clicks. The “Full” export button exports data, graphs and results together into a single file. Measurement data is stored in binary format and is thus fully tamperproof.

Theλ90/90 value is the basis for determination of the declared value of the thermal conductivity within the realm of CE declarations of building materials. It is calculated from a measurement series of at least 10 measurements and states which thermal conductivity values to a probability of 90%, can be achieved for 90% of the output production volume. The integrated report calculates the λ90/90 value by a mouse mouse click. The calculation is based on your measurements; no additional documentation and calculation are required.

Any measurement can be started by using the pre-defined instrument parameters (including number and position of thermocouples, stability criteria, etc.). However, experienced users who wish to apply their own parameter sets can define them under “Setup & Control”.

E-Learning

Become an Expert with our Free E-Learning Courses

All NETZSCH E-Learning Basic Courses are free of charge! The content is created by our laboratory method experts, who share their personal experiences with you. Take advantage of flexible online learning, fully adapted to your training needs!

Please accept Marketing Cookies to see that Video.

A red racing car is in a pit stop with crew members in red uniforms preparing for a fast tire change during a formula racing event.

Consultancy & Sales

Do you have further questions about the instrument or the method and would like to speak to a sales representative?

Service & Support

Do you already have an instrument and need technical support or spare parts?

AI Overview
An error occurred. Please try again.