STA 449 Nevio thermal analysis equipment with digital display, designed for precise sample measurement and analysis.
simultaneous thermogravimetry

STA 449 F3 Nevio

Specialized for Demanding Tasks in Research & Development 

The simultaneous thermal analyzer NETZSCH STA 449 F3  Nevio allows the measurement of mass changes and thermal effects under identical conditions at the same specimen.

Thus it makes it feasible to differentiate between e.g., structural changes and degradation phenomena within just one measurement up to 675°C1. The high flexibility caused by the various sensors, the great variety of sample crucibles and the wide TGA-measuring range make the system applicable for analysis of different kinds of pharmaceuticals, cosmetic samples or food including also inhomogenous substances. The vacuum-tight design of the instrument is an essential requirement for evolved gas analysis, i.e., the coupling of the thermal analyzer to a gas analyzing system such as FT-IR, MS or GC-MS.

Various furnaces, easy interchangeable by the user, cover a very broad temperature range (please see accessories). In addition, due to its modular setup the STA 449 F3 Nevio is also prepared to investigate samples under defined humid atmosphere. The analysis of measuring data runs under a standard PC and the innovative NETZSCH Proteus® software.

 

1⁾ optimized temperature range for pharmacy, cosmetics and foodstuffs; depending on the selected furnaces: total temperature range: -150°C up to 2400°C

Request a Quote

Get a personalized offer now.

Request a Quote

Technical Data

Temperature range
-120°C to 675°C ¹⁾
max. heating and cooling rates
50 K/min
max. measuring and weighing range
35000 mg ²⁾ /± 250 mW ³⁾
NETZSCH STA 449 Nevio thermal analyzer with a digital display, designed for high-precision testing and sample preparation.

Enthalpy accuracy:
± 1 ... 3 %

TG resolution:
0.1 µg (over entire weighing range)

Cooling options:
compressed air (<0°C to 675°C 1)), liquid nitrogen (-150°C to 675°C 1))

Atmospheres:
inert, oxidizing, reducing (forming gas), humid, vacuum, static, dynamic

Vacuum-tight:
Yes

1) optimized temperature range for pharmacy, cosmetics and foodstuffs; depending on the selected furnaces: total temperature range: -150°C up to 2400°C

2) minus weight of crucible

3) for thermocouple type E

Laboratory scene featuring scientists in white coats working with equipment, emphasizing quality and collaborative consulting for testing projects.

This instrument is LabV®®️-primed

LabV®®️ takes the data from your analytical instrument: It automatically imports all measurement data into a central and secure database solution, the LabV®®️ software. This allows you to visualize the data in LabV®®️ and to make them searchable. Your data will now be accessible from anywhere. Moreover you have the possibility to generate reports.

TypeTemperature rangeCooling system
Silver furnace-120°C to 675°Cliquid nitrogen
Copper furnace-150°C to 500°Cliquid nitrogen
Steel furnace -150°C to 1000°Cliquid nitrogen
Platinum furnaceRT to 1500°Cforced air
Silicon carbide furnaceRT to 1600°Cforced air
Rhodium furnaceRT to 1650°Cforced air
Graphite furnaceRT to 2000°Ctap or chilled water
Water vapor furnaceRT to 1250°Cforced air
High-speed furnaceRT to 1250°Cforced air
Tungsten furnaceRT to 2400°Ctap or chilled water

The STA 449 F3 Nevio runs under Proteus® Software on Windows®. Through the combination of easy-to-understand menus and automated routines, a tool has been created that is extremely user-friendly and, at the same time, allows sophisticated analysis. The Proteus® software is licensed with the instrument and can of course be installed on other computer systems.

 

Features concerning TGA plus DSC:

  • Proteus® Protect (option) - meeting the requirements of 21 CFR Part 11 and EU Annex 11
  • AutoEvaluation - autonomous and user-independent evaluation of thermal effects
  • Identify (option) - curve recognition and database system for library search purpose
  • SuperPosition - Simulation of the thermal behavior of physical mixtures by overlapping the curves of the mixture components incl. freely selectable mass ratios
  • Average curves - absolute (taking the sample masses of the various measurement s into account) or relative (applies just the number of measurements as denominator)
  • Report generator - based on a Microsoft Word plug-in each operator can easily create his own report template for reports – including logos, tables, description fields and plots.

 

TGA-specific evaluation features:

  • Mass changes in wt% or mg
  • Determination of residual mass
  • Extrapolated onset and end temperatures
  • Peak temperatures of the 1st and 2nd derivative of the mass-change curve
  • Automatic baseline and buoyancy correction
  • c-DTA® for calculated DTA signal with evaluation of characteristic temperatures and the peak area (optional for TGA measurements)

 

DSC-specific evaluation features:

Determination of the DSC conversion curve (in %) related to a partial peak area evaluation of a melting effect

  • TM-DSC module software extension for temperature-modulated DSC tests (optional)
  • Lissajous diagram for TM-DSC curves (optional)
  • Purity Determination (optional)

You can use the following software with this product

Different sample properties need different crucible materials/types. Therefore, a variety of sample crucibles (pans) made of alumina, platinum, aluminum, graphite, fused silica, etc. is available in different sizes.

Specific Furnaces for Specific Tasks

Due to its modular concept the STA can be equipped with standard furnaces for various temperature ranges but also with special furnaces - either alone or in combination by using a double hoisting device.

For TGA measurements under defined relative humidity the copper furnace plus the humidity generator can be applied. Regarding higher water concentrations (up to 100% water vapor) it is possible to use the water vapor furnace in combination with the water vapor generator.

The high speed furnace which can realize heating rates up to 1000 K/min is designed for TGA experiments as well.

Assorted Nevio STA 449 F3 accessories including lids and containers, showcasing precision components for testing and analysis.
STA 449 F3 Nevio accessories
Advanced laboratory equipment with a touchscreen monitor, set against a dark background with dynamic blue wave patterns.
Examples for STA 449 sample carrier and sensor types
A white laboratory connector featuring two cylindrical parts with a clasp for secure attachment, ideal for sample preparation.
High-quality STA measurement sensor with dual measurement heads designed for precise sample preparation analysis.

ASC - for Continous Operation

The automatic sample changer works day and night, giving you time for other tasks, and provides for optimal use of the STA 449 F3 Nevio (e.g. calibrations on the weekend).

A carousel is available for up to 20 sample and reference crucibles (pans) which can be automatically run in any order. The controlled generation of the required gas atmosphere in the sample chamber and cooling are automatic.

Of course, each sample can be assigned an individual measurement and evaluation procedure.

 

PulseTA®

With the PulseTA® technique an exactly defined quantity of gas is injected to the purge gas of the thermobalance (TGA) or the simultaneous thermal analysis instrument (STA).

This clearly increases the measuring possibilities:

  • Defined/incremental chemical reactions or adsorption of the injection gas with the sample
  • Quantification of gases evolved from the sample.
Schematic illustrating gas injection and analysis setup, featuring pulse device, TGA/DSC analysis, and gas detection systems.

Crucibles (Pans) Made of Aluminum for the STA 449 F3 Nevio

Material
(Purity)
Temperature
Range
Consisting ofDimension/
Volume
RemarksOrder Number
Al (99.5%)Max. 600°CCrucible + lidø 6 mm / 25/40 μlSet of 100, cold weldable*6.239.2-64.5.00
Al (99.5%)Max. 600°CCrucible + lidø 6 mm / 25/40 μlSet of 500, cold weldable*6.239.2-64.51.00
Al (99.5%)Max. 600°CCrucible + punched lid (0.05mm)ø 6 mm / 40 μlSet of 100, cold weldable*6.239.2-64.8.00
Al (99.5%)Max. 600°CCrucible + punched lid (0.05mm)ø 6 mm / 40 μlSet of 500, cold weldable*6.239.2-64.81.00
Al (99.5%)Max. 600°CCrucibleø 6.7 mm / 85 μlSet of 100 pieces,
non-weldable
NGB810405
Al (99.5%)Max. 600°CLid Set of 100 pieces, for
NGB810405
NGB810406

* Please note: One and the same sealing press is capable of sealing all the Al crucibles (pan + lid), order no. 6.240.10-80.0.00A

Product Information

We do not only promise quality, but also individual consulting. Let us start your project together.

Request Information