MMC 274 Nexus
Multiple Module Calorimeter

Multiple Module Calorimeter MMC 274 Nexus®

Various Modules in One Calorimeter

The Multiple Module Calorimeter (MMC)A multiple mode calorimeter device consisting of a base unit and exchangeable modules. One module is prepared for accelerating rate calorimetry (ARC), the ARC-Module. A second one is used for scanning tests (Scanning Module) and a third and fourth one is related to battery and polymer, pharmaceztical testing for coin cells (Coin Cell Module).MMC 274 Nexus® consists of two parts: the base unit with electronics and the exchangeable calorimeter module. This guarantees maximum flexibility.

Exchangeable calorimeter modules are available for different applications. The Accelerating Rate Calorimeter (Accelerating Rate Calorimetry (ARC)The method describing isothermal and adiabatic test procedures used to detect thermally exothermic decomposition reactions.ARC®) module can be used for standard process safety work. The External heater module can be employed for more sophisticated measurements.

NexSys MMC 274 device by NETZSCH, designed for material characterization and analysis, featuring a sleek black and white design.

Key Features and Advantages of MMC 274 Nexus®

Request a Quote

Get a personalized offer now.

Request a Quote

Technical Data

Temperature range
RT to 500°C
Temperature readability
0.01 K
Heating rate
0 to 5 K/min

The Multiple Module Calorimeter (MMC)A multiple mode calorimeter device consisting of a base unit and exchangeable modules. One module is prepared for accelerating rate calorimetry (ARC), the ARC-Module. A second one is used for scanning tests (Scanning Module) and a third and fourth one is related to battery and polymer, pharmaceztical testing for coin cells (Coin Cell Module).MMC 274 Nexus® can be used for a wide range of applications in different areas

Process Safety: The Accelerating Rate Calorimetry (ARC)The method describing isothermal and adiabatic test procedures used to detect thermally exothermic decomposition reactions.ARC® Module

Heat-Wait-Search (HWS)Heat-Wait-Search is a measurement mode used in calorimeter devices according to accelerating rate calorimetry (ARC).Heat-Wait-Search and Pressure Tests
A defined volume of a sample (ml scale) is placed in a tube-shaped container. The container is surrounded by a sophisticated heating system. Depending on the operating mode, the surroundings of the tube vessel are controlled to the same temperature as the sample. If there is no temperature difference between the surrounding heaters and the sample, then all the heat generated by the sample stays inside the sample.

Key Features of the ARC® Module

The Scanning ModuleA calorimeter module being part of the Multipe Module Calorimeter (MMC) allowing for scanning test of a sample. This procedure can serve as a screening test in order to detect a thermal hazard potential within a reasonably short measurement time.Scanning Module


IsothermalTests at controlled and constant temperature are called isothermal.Isothermal and Scanning Tests

Here, an additional heater is placed on the outside of the sample container. This allows for a defined power input and more accurate temperature control of the sample. The external heater is useful for running IsothermalTests at controlled and constant temperature are called isothermal.isothermal and constant temperature ramp tests, especially in those experiments where reaction energies are higher.

Key Features of the Calorimeter with the Scanning ModuleA calorimeter module being part of the Multipe Module Calorimeter (MMC) allowing for scanning test of a sample. This procedure can serve as a screening test in order to detect a thermal hazard potential within a reasonably short measurement time.Scanning Module

Stainless steel containers in various heights, designed for precision handling and storage in industrial applications.
Sample containers of different materials and volumes

Product Information

We do not only promise quality, but also individual consulting. Let us start your project together.

Request Information