Temperature range-180 to 2400°C
Heating rate0.001 K/min to 200 K/min
Max. sample weight/measuring range2 g (including crucible)
Gas atmospheresInert, oxidizing, reducing, measurements under vacuum (for tests such as rubber analysis)
Sample DimensionsCrucible volume up to 350 µl

Determination of the Glass Fiber Content in PA66

Polyamide 66 is a thermoplastic polymer which is used for a wide variety of technical parts.
The stiffness of the PA66 can be increased by suitable fillers such as glass fibers. TGA allows for determination of not only the Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition of the polymer, but also the precise glass fiber content (residual mass 20.2% at 989°C).
After switching from nitrogen to synthetic air, the TGA curve also shows the content of pyrolytic soot and Added Carbon BlackCarbon black acts as a reinforcing filler in tires and other rubber products. In other materials such as plastics, paints, and inks, carbon black is used as a color pigment or as filler to achieve electric conductivity.added carbon black. The c-DTA® peak exhibits the melting peak of PA66 at 260°C (red curve).

Thermogravimetric analysis (TGA) curve showing mass changes and thermal stability data for PA66 polymer under different atmospheres.