Laser Flash Technique (LFA)

based on, e.g., ASTM E1461, DIN EN 821-2, DIN 30905, ISO 22007-4, ISO 18755 

Principle of the LFA Method

The laser or light flash method is used for measurement of the Thermal DiffusivityThermal diffusivity (a with the unit mm2/s) is a material-specific property for characterizing unsteady heat conduction. This value describes how quickly a material reacts to a change in temperature.thermal diffusivity of a variety of different materials.

The front surface of a plane-parallel sample is heated by a light pulse and the resulting temperature rise at the sample’s rear face is recorded as a function of time. The higher the Thermal DiffusivityThermal diffusivity (a with the unit mm2/s) is a material-specific property for characterizing unsteady heat conduction. This value describes how quickly a material reacts to a change in temperature.thermal diffusivity, the faster the temperature rise reaches the rear face. 

Diagram illustrating the Laser Flash Technique (LFA) for thermal conductivity measurement, indicating sample position and equipment layout.
Figure 1
Graph illustrating the Laser Flash Technique (LFA) for thermal conductivity measurement, showing temperature rise over time.
Figure 2

In a one-dimensional, the Thermal DiffusivityThermal diffusivity (a with the unit mm2/s) is a material-specific property for characterizing unsteady heat conduction. This value describes how quickly a material reacts to a change in temperature.thermal diffusivity rise is calculated from this temperature rise as follows:

a = 0.1388 ⋅ d² /  t0.5 with
a – Thermal DiffusivityThermal diffusivity (a with the unit mm2/s) is a material-specific property for characterizing unsteady heat conduction. This value describes how quickly a material reacts to a change in temperature.Thermal diffusivity in cm²/s
d – Thickness of the sample in cm
t0.5 – Time to half minimum in s

With all NETZSCH laser flash apparatuses (LFAs), the Thermal DiffusivityThermal diffusivity (a with the unit mm2/s) is a material-specific property for characterizing unsteady heat conduction. This value describes how quickly a material reacts to a change in temperature.thermal diffusivity and Specific Heat Capacity (cp)Heat capacity is a material-specific physical quantity, determined by the amount of heat supplied to specimen, divided by the resulting temperature increase. The specific heat capacity is related to a unit mass of the specimen.specific heat capacity can be determined. This data is then used for calculation of the Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity.