| Published: 

Deeper Insight into the Pyrolysis of Acetylsalicylic Acid by Means of Thermogravimetric Measurements in Various Gas Atmospheres, Part 3

Introduction

In pharmacy, there is hardly any active ingredient about which more has been written than acetylsalicylic acid (or ASA for short; in English-speaking countries even the brand name Aspirin™ is often used as a synonym). Its success story began at the end of the 19th century when Dr. Felix Hoffmann synthesized the substance at the BAYER laboratories for the first time without impurities. Nowadays, it is still one of the most popular pharmaceuticals used across a broad therapeutic range. It belongs to the group of non-steroidal anti-inflammatory drugs (NSAIDs) and is indicated for the treatment of pain, fever and inflammation. In addition, it is used to prevent recurrence of heart attack or stroke in high-risk patients. In 1977, ASA was added as an analgesic to the “essential drug list” of the WHO (World Health Organization). [1] 

This is one of four application notes that examine in more detail the thermal behavior of acetylsalicylic acid: Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition in different gas atmospheres, Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition kinetics, and the resulting gas species.

Table 1: Measurement parameters

Parameter

Acetylsalicylic Acid

Sample mass

Approx. 5 mg

Atmosphere

Argon, nitrogen and helium

Crucible

Al2O3, 85 μl, open

Temperature program

RT to 600°C

Flow rate

40 ml/min

Sample holder

TGA, Type P

Experimental

The sample material, acetylsalicylic acid (CAS: 50-78-2), was acquired from Sigma Aldrich with a purity of > 99%. It is a white, crystalline powder which exists in three crystal modifications [2]. Form I, with a Melting Temperatures and EnthalpiesThe enthalpy of fusion of a substance, also known as latent heat, is a measure of the energy input, typically heat, which is necessary to convert a substance from solid to liquid state. The melting point of a substance is the temperature at which it changes state from solid (crystalline) to liquid (isotropic melt).melting point of about 137°C [4], is the most stable one at ambient temperature and above [3]. 

PyrolysisPyrolysis is the thermal decomposition of organic compounds in an inert atmosphere.Pyrolysis experiments are usually carried out in a nitrogen atmosphere due to its availability and comparatively low price. This is also reflected in several publications, for example [5] and [6]. To answer the question as to wheter the results obtained under nitrogen can also be generalized to other atmospheres, an experimental series was performed for studying the thermal behavior of acetylsalicylic acid as a function of the purge gas nature. Besides nitrogen, other inert gas atmospheres used were helium and argon. The corresponding measurement parameters are summarized in table 1.

For characterization of the thermal behavior, a NETZSCH TG 209 F1 instrument was employed.

Figure 1 depicts the typical two-step Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition profile of acetylsalicylic acid, resulting from heating the substance in a flowing argon atmosphere. The first step, with a DTG peak temperature of about 190°C, exhibits a mass loss of 47%; the second step, at 361°C (again DTG peak temperature), exhibits almost 53%. However, no real plateau occurs between the two mass-loss steps. The first one merges, more or less, into the second. This indicates that maybe more than two Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition steps are involved. The possibility that a more complex mechanism such as this would be the case here is additionally supported by the fact that the second DTG peak has a clearly visible shoulder at approx. 320°C in the descending slope.

Comparing the situation in an argon atmosphere with experiments performed at identical heating rates in both nitrogen and helium atmospheres (figure 2), the behavior under nitrogen conditions is about the same as in argon, whereas it changes significantly under a flowing helium atmosphere. An increase in the first massloss step of about 18 percentage points (from 47% to almost 65%) and consequently, a decrease in the second mass-loss step of the same amount (from 53% to 35%) is observed. In addition, both mass-loss steps are shifted to somewhat lower temperatures, indicated by the decrease in the particular DTG peak temperatures (4 K to 7 K for the first DTG peak and 11 K to 15 K for the second one). This suggests that something different is going on in a helium atmosphere than in argon and nitrogen atmospheres.

Thermogravimetric analysis (TGA) and derivative thermogravimetry (DTG) of acetylsalicylic acid in argon, showing mass loss percentages across temperatures.
1) Temperature-dependent mass loss (TGA) and mass-loss rate (DTG) of acetylsalicylic acid in an argon atmosphere; sample mass: 5.02 mg; open Al2O3 crucible
TGA and DTG curves of acetylsalicylic acid in argon, nitrogen, and helium atmospheres, showing temperature-dependent mass loss.
2) Temperature-dependent mass loss (TGA) and mass-loss rate (DTG) of acetylsalicylic acid in an argon (red, solid line), nitrogen (black, dashed line) and helium atmosphere (blue, solid line); open Al2O3 crucible

 

In literature, a two-step mechanism with simultaneous evaporation of intermediates is proposed as the PyrolysisPyrolysis is the thermal decomposition of organic compounds in an inert atmosphere.pyrolysis mechanism of acetylsalicylic acid [7]. Measurements conducted by NETZSCH in a nitrogen as well as in a helium atmosphere using hyphenated techniques, more precisely, TGA/STA in combination with FT-IR [6] and GC-MS [8], support this hypothesis. This indicates that the nature of the Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition products is independent of the gas atmosphere. 

The major difference between all these experiments is the DensityThe mass density is defined as the ratio between mass and volume. density of the purge gas used (see table 2). It differs maximally by a factor of 10. 

This suggests that a higher DensityThe mass density is defined as the ratio between mass and volume. density of the purge gas creates a higher back pressure, which results in a reduced transfer of the volatile sample components into the gas atmosphere. This effect is particularly visible when using helium, which has a much lower DensityThe mass density is defined as the ratio between mass and volume. density than nitrogen or argon. Since true Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition reactions are independent of the surrounding inert gas atmosphere [10], it is perhaps the parallel evaporation that is mostly affected. 

The fact that the decomposition takes place at slightly lower temperatures in helium (e.g., DTG peak at 183°C compared to 187°C in nitrogen and approx. 190°C in argon) is due to the higher Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity of this gas (see table 3). In the temperature range where thermal radiation plays only a minor role, the sample reaches the reaction temperature in a purge gas with higher Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity somewhat earlier.


Table 2: DensityThe mass density is defined as the ratio between mass and volume. Density values at 0°C and normal pressure of various purge gases

Gas

DensityThe mass density is defined as the ratio between mass and volume. Density / (g/cm³) [9]

Helium

0.178

Nitrogen

1.251

Argon

1.784


Table 3: Thermal conductivity values under standard conditions of the various purge gases

Gas

Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.Thermal Conductivity (W/m·K) [11]

Helium

0.1567

Nitrogen

0.0260

Argon

0.0179

Conclusion

The present example shows that the selected gas atmosphere may have a strong impact on the thermogravimetric measurement results, even if the purge gas does not act as a reaction partner. A greatly varying gas DensityThe mass density is defined as the ratio between mass and volume. density can have an effect on the transfer of gaseous compounds from the sample surface into the surrounding gas atmosphere – especially if evaporation is involved.

Literature

  1. [1]
  2. [2]
    A.G. Shtukenberg, C.T. Hu, Q. Zhu, M.U. Schmidt,W. Xu, M. Tan and B. Kahr, Cryst. Growth Des., 17, 2017,pp 3562–3566
  3. [3]
  4. [4]
    P.P. Bag and C.M. Reddy, Cryst Growth Des., 12, 2012,pp 2740 - 2743
  5. [5]
    E. Füglein et al., NETZSCH Application Note 135
  6. [6]
    Y.A. Ribeiro, A.C.F. Caires, N. Boralle, M. Ionashiro,Thermochim Acta 279, 1996, pp 177-181
  7. [7]
    C. Fischer, NETZSCH Application Note 208
  8. [8]
  9. [9]
    Measurements on calcium oxalate monohydrate,carried out in the NETZSCH applications laboratory
  10. [10]
AI Overview
An error occurred. Please try again.