Thermal Conductivity of Inhomogeneous Refractories − LFA Versus Hot Wire Method

Introduction

Thermophysical properties such as Thermal DiffusivityThermal diffusivity (a with the unit mm2/s) is a material-specific property for characterizing unsteady heat conduction. This value describes how quickly a material reacts to a change in temperature.thermal diffusivity, specific heat, and Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity are crucial parameters for optimizing the production and application of coarse ceramics. For decades, such properties have been determined from stationary methods (e.g., guarded hot plate technique) or standardized transient techniques such as the hot-wire method according to ISO 8894 (see TCT 426 in figure 1). However, these methods are limited to large sample sizes and low thermal conductivities. Moreover, these methods are very time-consuming. 

Flash methods are non-contact measurement techniques and can handle high Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity materials without any difficulties. Additionally, flash methods are absolute methods for determining Thermal DiffusivityThermal diffusivity (a with the unit mm2/s) is a material-specific property for characterizing unsteady heat conduction. This value describes how quickly a material reacts to a change in temperature.thermal diffusivity. Modern instruments (see LFA 427 in figure 2) often also allow the simultaneous measurement of the specific heat of a material so that Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity can be determined without additional measurements. Since flash methods are limited to homogeneous samples with small dimensions, these methods have not been used for analyzing inhomogeneous coarse ceramics. However, using state-of-the-art highly sensitive systems, testing of larger samples becomes possible now [1]. In addition, the fast testing times of the flash methods allow for measurements of various samples taken from the brick to be tested without further efforts. In the work described herein, the results from laser flash and the hot-wire measurements on a silicon carbide-containing brick and a magnesia-spinel brick are compared. Measurements were carried out on several small samples of the same material to check the homogeneity of the material and the reproducibility of the methods.

NETZSCH TCT 426 testing equipment with open door, showcasing stacked test materials for thermal analysis and testing.
1) NETZSCH TCT 426
NETZSCH LFA 427 laser flash analyzer for thermal conductivity testing, featuring a sleek design and advanced measurement controls.
2) NETZSCH LFA 427

Test Results

Figure 3 depicts the results of Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity measurements on a magnesia-spinel brick (figure 4) using the LFA 427 and the TCT 426. The dashed line represents the average values (error bar ±10%) of combined data from the two different methods. It can clearly be seen that most values of independent LFA and TCT measurements are within the ±10% range from the average. This illustrates the high accuracy of both systems.

Graph comparing thermal conductivity of magnesia-spinel bricks at various temperatures; shows LFA and TCT test results with error bars.
3) Thermal conductivity of the magnesia-spinel brick; comparison of LFA and TCT test results; 10% error bar for the average values of all TCT/LFA results.
Magnesia-spinel brick alongside three graphite-coated LFA samples, showing testing materials for thermal conductivity analysis.
4) Magnesia-spinel brick (85% MgO, and 12% Al2O3), measured with the TCT 426 (left) and the LFA 427 (right); the 3 LFA samples were coated with graphite.

Furthermore, the deviation between the different samples shows the possible range of the Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity due to the inhomogeneity of the magnesia-spinel brick. A similar comparison of LFA and TCT measurements on silicon carbide-containing brick (figure 6) is shown in figure 5. Again, the independent measurement values are all within ±10% of the averaged data from the two methods combined.

Thermal conductivity analysis of silicon carbide bricks, displaying LFA and TCT results against temperature with error bars.
5) Thermal conductivity of silicon carbide containing brick; comparison of LFA and TCT results; 10% error bar for the average values of all TCT/LFA results.
Silicone carbide brick and three LFA samples, with graphite coating, used for thermal conductivity testing.
6) Silicone carbide-containing brick (45% Al2O3, 29% SiO2 and 25% Sic), measured with the TCT 426 (left) and LFA 427 (right); the 3 LFA samples were coated with graphite.

Conclusion

The good agreement between the results obtained by the two different methods laser flash and hot wire clearly show that both methods are very suitable for analyzing refractories with a high accuracy. The NETZSCH model LFA 427, however, offers several advantages. Test results can be achieved quickly and with a high accuracy. The speed of measurement compensates for the small sample dimensions as more samples can be tested with a higher sample throughput. TCT measurements are much more time-consuming due to the huge sample dimension and the long stabilization time. However, the hot-wire method according to ISO 8894 is highly demanded for refractory materials.

Literature

  1. [1]
    J. Blumm, A. Lindemann: Thermophysical Properties Characterization of Coarse Ceramics Using the Transient Laser Flash Technique. J.G. Heinrich and and Aneziris, Proc 10th ECerS Conf., Göller Verlag, Baden-Baden,2007, 205-211, ISBN: 3-87264-022.4