NETZSCH STA 319 Jupiter simultaneous thermal analyzer with touchscreen, automatic sample changer, and gas analysis integration.
DSC-TGA analysis

STA 319 Jupiter®

Offering Access to Simultaneous Thermal Analysis

Highlights

Experience the next generation of simultaneous thermal analyzers with the STA 319 Jupiter®

The STA 319 Jupiter® is part of NETZSCH's Jupiter® range of STAs and offers several advantages:

  • Temperature range of 10°C (RT) to 1025°C (or up to 1100°C): ideal for polymer, life science and inorganic measurements.
  • Robust water-cooled ceramic furnace for fast heating and cooling.
  • Advanced top-loading microbalance: Provides an impressive 20-ng resolution for long-term, high-quality measurements.
  • Revolutionary ring sample holder design: eliminates the need for a standard reference crucible. Enables quick setup and minimizes handling errors.
  • Defined atmospheric conditions thanks to vacuum-tight design.
  • Coupling to gas analysis enables real-time identification of the composition of evolved gases.

Check Out all Details of the STA 319 Jupiter®

Please accept Marketing Cookies to see that Video.

Discover precise, flexible simultaneous thermal analysis with rapid temperature control, 20 ng resolution, controlled atmospheres, and real-time evolved gas analysis.
Thin film ribbon cable thermal conductivity testing
Magnetic Shock Absorbers

An optional magnetic levitation device can be used to elevate the instrument off the work bench. This eliminates the effect of external factors such as vibrations and contact, thus decreasing the number of measurements needed to produce reliable results.

Method

Thermogravimetric Analysis (TGA) / Differential Scanning Calorimetry (DSC)

What is Simultaneous Thermal Analysis?

Simultaneous Thermal Analysis (STA) is a method used to analyze the Thermal StabilityA material is thermally stable if it does not decompose under the influence of temperature. One way to determine the thermal stability of a substance is to use a TGA (thermogravimetric analyzer). thermal stability and composition of materials. It combines two techniques: Thermogravimetric Analysis (TGA) and Differential Scanning Calorimetry (DSC).

TGA measures the weight change of a material when it is heated or cooled, providing information on Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition temperatures, moisture content, and Thermal StabilityA material is thermally stable if it does not decompose under the influence of temperature. One way to determine the thermal stability of a substance is to use a TGA (thermogravimetric analyzer). thermal stability. DSC measures the amount of heat required to increase the temperature of a sample, providing insights into Phase TransitionsThe term phase transition (or phase change) is most commonly used to describe transitions between the solid, liquid and gaseous states.phase transitions, heat capacity, and reaction enthalpies.

STA combines these two measurement methods by simultaneously recording weight change and heat flow of a sample. A small amount of the material is placed in a special container and is then heated or cooled. The STA instrument records both the weight changes and the heat absorbed or released by the sample. This method is particularly useful because it provides detailed data on the thermal properties and composition of a material, and its efficiency saves time and sample material.

STA is used in quality control, research and development, and failure analysis. It is essential for studying the behavior of polymers, pharmaceuticals, food, and other materials under different temperature conditions, making it an invaluable tool for scientists and engineers.

Vertical top-loading TGA design diagram showing sample furnace, thermobalance, and components for thermal analysis.

Measuring Principle

Here's how Simultaneous Thermal Analysis works:

  1. Sample Preparation: A small amount of the material (sample) is placed in a special container (crucible).
  2. Heating/Cooling: The sample is heated or cooled at a controlled rate.
  3. Measurement: As the temperature changes, the STA instrument records:
    1. Weight change: How much the sample's weight increases or decreases.
    2. Heat flow: How much heat is absorbed or released by the sample.

The STA Masterclass:
Your Monthly STA Webinar Series

Our webinar series brings you practical, application-first sessions on STA - from phase changes to OxidationOxidation can describe different processes in the context of thermal analysis.oxidation. Learn directly from NETZSCH experts and enhance your daily analysis work.

Explore All Webinars

Specifications

Technical Data

Temperature range
RT (10°C) to 1025 / 1100°C at the sample
Balance resolution
20 ng
Heating and cooling rates
0.001 K/min to 300 K/min
Gas Outlet
Furnace
Furnace Water Cooling
Purge gas sample
Protective gas balance chamber
Balance
Sample holder lifting device
Sample Carrier

Fast cooling time:
In nitrogen: ≈ 12 min from 1100°C to 100°C
In helium: ≈ 5 min from 1100°C to 100°C, ≈10 min to 25°C

Wide measuring range:
2000 mg

Vacuum-tightness:
<< 10-1 mbar

Sample crucible volume:
up to 300 µl

TGA accuracy:
0.0003%

DSC accuracy:
± 5%

STA 319 Jupiter thermal analyzer with touchscreen displaying simultaneous TGA-DSC data, advanced sample handling by NETZSCH.
Laboratory workspace featuring organized vials in a container, ideal for research and sample analysis.

High-End 204-Sample ASC and Removable Sample Trays

The STA 319 Jupiter® ASC is designed to hold two interchangeable sample trays in microplate format, each holding 96 samples. This allows for clear assignment of the samples when they are prepared away from the instrument. An additional fixed strip is reserved for up to 12 calibration materials or empty crucibles of different dimensions and materials for calibration and correction purposes. Of course, correction measurements with empty crucibles can also be defined on the trays.

Advanced thermal analysis device NETZSCH TC 309 Libra, featuring a touch screen displaying graph data.
LED Status Bar
Color Touch Display

Software

STA 319 Jupiter® with Proteus® - Our Powerful Analytical Software

TGA and DTG curves representing the thermal analysis of a styrene-based thermoplastic elastomer, highlighting key temperature points and weight loss percentages.
STA measurement on CuSO₄·5H₂O (13.72 mg) applying a heating rate of 10 K/min, N2 atmosphere (70 ml/min) and PtRh crucibles with pierced lids; the DTG curve (black) was created by a mouse click and all evaluations were performed by AutoEvaluation

AutoEvaluation

Objective Results Right After a Measurement

AutoEvaluation is the industry’s first fully self-operating evaluation system for TGA and DSC measurements, setting a new benchmark in thermal analysis. It automatically detects and analyzes key events, including mass changes, EndothermicA sample transition or a reaction is endothermic if heat is needed for the conversion.endothermic and ExothermicA sample transition or a reaction is exothermic if heat is generated.exothermic reactions, DTG curves, and peak temperatures, without requiring user input. Results are displayed in real time after each measurement, and there are flexible options to customize detection settings and visualization. AutoEvaluation empowers both beginners and experienced professionals to achieve faster, more reliable results with ease by combining speed, accuracy, and unbiased analysis.

Identify

The Database for Material Identification and Quality Control

Identify is a powerful, unique software solution for identifying and classifying materials in the field of thermal analysis. With over 1,300 reference entries spanning polymers, organics, pharmaceuticals, food, cosmetics, inorganics, ceramics, metals, and alloys from the extensive NETZSCH libraries, it delivers comprehensive coverage across industries. Identify supports a variety of signal types, including DSC,DSC-Specific Heat Capacity (cp)Heat capacity is a material-specific physical quantity, determined by the amount of heat supplied to specimen, divided by the resulting temperature increase. The specific heat capacity is related to a unit mass of the specimen.cp, TGA, TGA-c-DTA®®, STA, DIL/TMA, and DMA. Users can easily expand the database with their own measurement data, creating a personalized library of unlimited size. With each new entry, Identify becomes an even more valuable tool, helping users compare results, interpret data, and plan future experiments with greater confidence and speed.

TGA and DSC analysis curves for an unknown material, showing temperature peaks and comparison data for identification.

It is particularly advantageous that Identify can even simultaneously incorporate two types of measurements, such as TGA and DSC or c-DTA®, during identification. As shown in the example above, analysis with Identify in the temperature range below 500°C reveals that the TGA-DSC results are very similar to those found for gypsum (dihydrate,CaSO4·2H2O) in the database. The DSC peak detected at 575°C, which is due to the structural α→β transition of quartz, occurs also in the most similar database curve in that temperature range. Above 600°C, the best hit from the database search is a measurement showing the Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition of calcium carbonate. In summary, the investigation showed that the unknown material consists of gypsum, quartz and calcium carbonate.

Sunlight filters through dense green foliage in a tranquil forest, creating a serene and inviting atmosphere.

Eco Mode – Uses Energy Only when Needed

FOCUS ON EFFICENCY AND A MORE SUSTAINABLE LABORATORY

Promote sustainability without compromising performance. Modern laboratories must balance high-quality research with responsible use of resources, an area in which our solution excels.

Eco Mode transforms the way STA instruments operate. Rather than wasting energy during idle periods or experiencing long warm-up times, this intelligent, software-controlled feature automatically powers down the chiller when not in use. It then reactivates gas flow and temperature control based on your schedule. Your instrument is always ready when you need it, with no unnecessary downtime or energy loss.

You can reduce power consumption by over 700 watts during inactive phases, save up to 1,800 kWh annually, and cut gas usage by up to 50%. The result is lower operating costs, improved efficiency, and a substantially reduced carbon footprint, helping you create a more sustainable, future-ready laboratory.

Our Quality Promise:

NETZSCH’s Unlimited Warranty

At NETZSCH, our commitment to quality goes beyond the instruments themselves. We understand that your investment in advanced technology is a long-term one, and and that’s why we offer something truly unique – our Unlimited Warranty.

3D mass-loss curve and FT-IR spectra of gases released during DCP heating, showcasing absorbance and temperature variations.

This instrument is LabV®️-ready.

Turn data into better products with LabV®—a centralized, AI-driven material intelligence platform that integrates data from this instrument and any other data sources. Designed for R&D and QC engineers, LabV® enables data-driven decision-making to drive innovation and ensure consistently high-quality results.

E-Learning

Become an Expert with our Free E-Learning Courses

All NETZSCH E-Learning Basic Courses are free of charge! The content is created by our laboratory method experts, who share their personal experiences with you. Take advantage of flexible online learning, fully adapted to your training needs!

Please accept Marketing Cookies to see that Video.

A Formula 1 car in the pit, surrounded by team members in red uniforms, preparing for a quick tire change during a race.

Consultancy & Sales

Do you have further questions about the instrument or the method and would like to speak to a sales representative?

Service & Support

Do you already have an instrument and need technical support or spare parts?

Related Devices

  • STA 509 Jupiter Classic
    STA 509 Jupiter®Classic

    Best Price/Performance Ratio

    • RT to 1600°C
    • SiC furnace
    • Balance resolution: 0.1 μg
    • Optional 20-position ASC
  • STA 509 Jupiter Select
    STA 509 Jupiter®Select

    Tailored to Your Needs

    • -150 to 2400°C
    • Choice of 12 different furnaces
    • Balance resolution: 0.1 μg
    • Optional 20-position ASC or2nd furnace
  • STA 509 Jupiter Supreme
    STA 509 Jupiter®Supreme

    Instrument for Highest Performance

    • -150°C to 2000°C
    • Choice of 9 different furnaces
    • Balance resolution: 0.025 μg
    • Optional 20-position ASC or2nd furnace