はじめに
Diamond powder combines exceptional hardness and wear resistance with outstanding thermal transport properties. It is therefore widely used for grinding and polishing hard materials such as cemented carbides, ceramics, gemstones and optical glass. In thermal interface materials, including thermal greases and pastes, diamond powder can also serve as a highly conductive filler, to help transfer heat from electronic components to heat sinks and thereby reduce operating temperatures. Reliable characterization of its thermal transport behaviour is thus essential for material development and quality control.
計測の課題
The thermal diffusivity and conductivity of a powder are not only dominated by the intrinsic properties of the material, but also by the particle size, packing density or the degree of compaction. Contact resistance between adjacent particles and air-filled voids within the powder bed can significantly reduce the effective thermal conductivity compared to that of bulk diamond.
Reproducible sample preparation and a measurement configuration suited to powders are therefore critical for meaningful comparative results.
Measurement with the LFA 717 HyperFlash®
The LFA results enables the temperature-dependent thermal diffusivity of diamond powder to be determined using either a sapphire sample holder or a pressure sample holder (see https://analyzing-testing.NETZSCH.com/en/products/thermal-diffusivity-and-conductivity).
The thermal conductivity can subsequently be calculated from the thermal diffusivity, density and specific heat capacity, with the latter determined by Differential Scanning Calorimetry (DSC). The pressure sample holder additionally allows the influence of different compaction pressures on thermal transport to be investigated.
In measurement 1, the diamond powder was analyzed in a sapphire sample holder. As can be seen, when the sample is placed naturally (without additional pressure) in the holder, its thermal diffusivity is approximately 0.4 mm²/s (Table 1).
Thermal diffusivity results for the diamond powder sample, measured in a sapphire sample holder. Table 1
表1:サファイア製試料ホルダーを用いて測定した、ダイヤモンド粉末試料の熱拡散率の結果です。

試料ホルダーに充填した後も粉末の表面が粗いままであったため、測定された温度上昇曲線については、浸透モデル*の方がより適切に説明できました(図1)。

* 標準モデルでは、パルスエネルギーが試料の前面で完全に吸収されると仮定しています。しかし、わずかに多孔質な材料の場合、吸収は前面に限定されず、試料内部の薄い層にまで及びます。 この吸収層の厚さは、材料内における光子の平均自由行程を表すものと見なすことができ、その結果、試料内部に向かって指数関数的に減衰する初期温度分布が得られます。このモデルは、1999年にMcMastersらによって提唱された手法を取り入れ、さらに改良を加えたものです[1]。
測定2では、ダイヤモンド粉末を圧力試料ホルダー内で分析しました(図2)。上下の支持板はサファイア製であったため、単層モデルを採用しました。トルクは40 cN・mに設定し、これは約0.5 MPaの圧力に相当します。

粉末試料に所定の圧力を加えたところ、粉末の充填密度が高まったことにより熱拡散率が上昇し、約1.2 mm²/sに達しました(表2)。また、熱拡散率を決定するために、浸透モデルも適用されました。
表2:加圧試料ホルダー内で測定されたダイヤモンド粉末試料の熱拡散率の結果

Results and Discussion
Thermal diffusivity results for the diamond powder sample, measured in a pressure sample holder. Table 2
When a defined pressure was applied to the powder sample, the thermal diffusivity increased due to the more compact packing of the powder, reaching approximately 1.2 mm²/s (Table 2). The penetration model was also applied to determine the thermal diffusivity.
The measured thermal diffusivity of the diamond powder is significantly lower than the values typically observed for bulk diamond (see Application Notes 438 and 476). This difference can be attributed to the pronounced thermal contact resistance between the individual particles and the insulating effect of air voids within the powder bed. Compared with compact solids, measurements on powder samples are generally expected to exhibit greater scatter. This is because the heat transfer in powders is strongly influenced by the structure and morphology of the powder bed. Relevant factors include particle size, particle shape, particle size distribution, and packing density. Together, these parameters determine the particle-to-particle contact and thus affect the effective heat transport through the powder bed. Consequently, variations in powder morphology and packing can have a significant impact on the measured thermal diffusivity. These results highlight the importance of considering powder morphology and compaction conditions when evaluating diamond-filled thermal interface materials.
Conclusion
The LFA 717 HyperFlash® offers a flexible approach for characterizing the thermal diffusivity of diamond powders. Dedicated sample holders accommodate loose or compacted powders, while suitable evaluation models account for sample-specific effects such as a rough surface. Combined with density and DSC-derived specific heat capacity, the measurement delivers the data required to determine thermal conductivity and assess diamond powders for demanding thermal-management applications.