Published on 

Thermal Characterization of Porous Concrete - Influence of Density on Thermal Conductivity Using the Heat-Flow Meter and Guarded Hot Plate Method

Advanced DSC 300 Caliris Classic thermal analysis instrument, featured with a user-friendly touchscreen and precise measurement capabilities.
1) GHP 456 HT Titan®
HFM 446 Lambda Medium, a compact analytical device for testing and analysis, featuring a user-friendly interface and sleek design.
2) HFM 446 LambdaMedium

Measurement Parameters

The effective Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity of porous materials depends heavily on the DensityThe mass density is defined as the ratio between mass and volume. density. Two specimens of porous concrete (300 mm x 250 mm x 60 mm, see figure 3) with slightly different densities were examined with regard to their Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity using the HFM 446 LambdaMedium (figure 2) and the GHP 456 HT Titan® (figure 1) from 10°C to 75°C. 

The HFM 446 LambdaMedium applies a relative method with an asymmetric setup using a calibration of heatflux sensors with a known reference material. The specimens are investigated individually. The GHP 456 HT Titan® applies an absolute method with a symmetric setup which uses two similar specimens for the measurement.

Porous concrete specimen showcasing a textured surface and detailed structure to explore material properties.
3) Porous concrete – Specimen 1 (left); porous structure n a breading edge (right)

Measurement Results

In the present case, the DensityThe mass density is defined as the ratio between mass and volume. density of the two porous concrete specimens was slightly different. Specimen 1 had a DensityThe mass density is defined as the ratio between mass and volume. density of approx. 516 kg/m³ and Specimen 2 of approx. 543 kg/m³ (difference ~5%). Figure 4 depicts the Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity of both concrete specimens. The orange points represent the measurement values for Specimen 1 measured with the HFM; the blue points represent Specimen 2. Specimen 1 shows a 6-7% lower Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity than Specimen 2. The mean values calculated using the individual measurements from the HFM coincide almost perfectly with the values from the GHP measurement where both specimens were used. The deviation is less than 0.8%.

Comparison of thermal conductivity in W/(m·K) for two specimens and GHP across varying temperatures.
4) Comparison of the Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity measured with the GHP and HFM.

The steady-state method for determining Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity requires a constant and one-dimensional heat flow through the specimens at all time. This is realized by continuously applying a heat source and sink to the specimen. 
Measurement technique: Heat Flow Meter (HFM) and Guarded Hot Plate Apparatus (GHP) 

In contrast, in transient methods, the heat energy transferred through the specimen is not constant. The heat flow rate varies. This can be caused, for example, by a short energy impulse on the specimen. Measurement techniques: Laser (Light) Flash Analysis (LFA)

Conclusion

The Thermal ConductivityThermal conductivity (λ with the unit W/(m•K)) describes the transport of energy – in the form of heat – through a body of mass as the result of a temperature gradient (see fig. 1). According to the second law of thermodynamics, heat always flows in the direction of the lower temperature.thermal conductivity of two different porous concrete specimens was investigated with two different steady state methods. The HFM measurements on the individual specimens show the differences caused by the different densities of the specimens. Also, the GHP device can handle samples with slightly different properties yielding an appropriate mean value. Both instruments are well suited for the characterization of insulating materials.

AI Overview
An error occurred. Please try again.