The heated gas cell can be located in an external module or connected to the INVENIO via the right exit, or the gas cell may be situated at the internal position.
An external gas cell allows the operator access to the FT-IR spectrometer’s great flexibility and makes it possible to add on accessories such as an ATR unit or a microscope.
The gas transfer line is characterized by a low volume and a short capillary length, thus minimizing dilution effects and leading to fast response times. To prevent cold spots and thus condensation of the gases evolved, the entire gas path is heated.

Request a Quote
Technical Data
Gas cell volume and length
Temperature of transfer line
Detector

Wave number range:
FT-IR: 8000 cm-1 to 340 cm-1
Coupling: 4400 cm-1 to 600 cm-1
(double window technique KBr+ZnSe)
Resolution:
better than 0.4 cm-1
Furnace adaptor:
max. 400°C
Transfer line material:
Stainless steel (exchangeable)
Window material gas cell:
ZnSe + KBr
Application Literature
- PTFE – A Fascinating Polymer Investigated by Advanced Thermal Analysis TechniquesApplication Note
- Material Characterization of Polymers by Methods of Thermal AnalysisApplication Note
- Identification of Polymers in Blends by Means of TGA FT-IR and the NETZSCH-FT-IR Database of PolymerApplication Note
- Simultaneous Detection of Organic Functional Groups and Inorganic Molecules in Thiol-Capped ZnO Nanoparticles with Evolved Gas Analysis by FT-IR and MSApplication Note
- TGA-FT-IR Coupling – Improved Sensitivity by Means of Fast Heating RatesApplication Note
- TGA-GC-MS Measurements on Pinewood/SapwoodApplication Note
- Polyvynil ChlorideApplication Sheet
- Polycarbonate (PC)Application Sheet