Published on 

Deeper Insight Into the Pyrolysis of Acetylsalicylic Acid by Means of Thermogravimetry and GC-MS, Part 1

Introduction

In pharmacy, there is hardly any active ingredient about which more has been written than acetylsalicylic acid (or ASA for short; in English-speaking countries even the brand name Aspirin™ is often used as a synonym). Its success story began at the end of the 19th century when Dr. Felix Hoffmann synthesized the substance at the BAYER laboratories for the first time without impurities. Nowadays, it is still one of the most popular pharmaceuticals used across a broad therapeutic range. It belongs to the group of non-steroidal anti-inflammatory drugs (NSAIDs) and is indicated for the treatment of pain, fever and inflammation. In addition, it is used to prevent recurrence of heart attack or stroke in high-risk patients. In 1977, ASA was added as an analgesic to the “essential drug list” of the WHO (World Health Organization). [1] 

This is one of four application notes that examine in more detail the thermal behavior of acetylsalicylic acid: Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. Decomposition in different gas atmospheres, Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition kinetics, and the resulting gas species. [2, 3, 4]


Table 1: STA measurement parameters

Parameter

Acetylsalicylic Acid

Sample mass

4.96 mg

Atmosphere

Helium

Crucible

Al2O3, 85 μl, open

Temperature program

RT to 50 °C, 10 K/min

Flow rate

100 ml/min

Sample holder

TGA, Type S

Results and Discussion

For the investigation of the thermal Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition of acetylsalicylic acid, thermogravimetric measurements (TGA) were carried out with the NETZSCH STA 449 F3 Jupiter®, coupled to a GC-MS system (Agilent 8890 gas chromatograph and Agilent 5975 MSD). Inert gases such as helium were used as the purge gas atmosphere. Detailed information about the measurement conditions is summarized in table 1.

The PyrolysisPyrolysis is the thermal decomposition of organic compounds in an inert atmosphere.pyrolysis of acetylsalicylic acid exhibits two mass-loss steps (see figure 1). The first mass loss of 66.4% is related to a peak in the mass-loss rate (DTG) at 170°C. The second mass-loss step amounts to 33.4% with a peak in the DTG curve at 327°C.

In order to provide insight into the PyrolysisPyrolysis is the thermal decomposition of organic compounds in an inert atmosphere.pyrolysis products, the TGA-GC-MS coupling was used to separate the complex mixture of gases and to identify the different components. The measurement parameters for GC-MS are described in table 2.

TGA and DTG graphs showing temperature-dependent mass loss of acetylsalicylic acid in helium, highlighting key temperatures.
1) Temperature-dependent mass loss (TGA) and mass-loss rate (DTG) of acetylsalicylic acidin a helium atmosphere

 

Table 2: GC-MS measurement parameters

Parameter

Cryo Trapping Mode

Column

Agilent HP-5ms

Column length

30 m

Column diameter

0.25 mm

Cryo trap

-50°C, 45 min

Column temperature

40°C, isotherm, 48 min

40°C to 300°C, 15 K/min

Gas atmosphere

He

Column flow (split)

2 ml/min (5:1)

Valve

Every 1 min

The gases released were sampled every minute on the cryo trap. After the thermogravimetric run, the cryo trap was heated from -50°C to 300°C at a heating rate of 300 K/min to vaporize the condensed compounds and let them separate over the GC column (which was heated at 15 K/min). This method increases the concentration of side products and enables excellent separation. The resulting total ion current is displayed in figure 2. Comparison of the detected MS spectra for each peak with the NIST library yields a number of compounds with excellent hit quality. Identification examples are shown for the peaks with a retention time of 59.31 min and 60.89 min in figures 3 and 4. Besides acetic acid, phenol, salicylic acid, and acetylsalicylic acid, cyclic oligomers of 2-hydroxy benzoic acid as stated in literature were also found. This analysis reveals that Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition and evaporation take place simultaneously and furthermore explains why the two mass-loss steps are not separated.

Chromatogram displaying evolved decomposition products of acetylsalicylic acid over time, highlighting key compound peaks.
2) Chromatogram of the evolved Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition products of acetylsalicylic acid collected by the cryo trap

 

Library Search

Retention time [min]

Name

Hit quality
49.89

Acetic acid

91
55.58

Phenol

96
56.63

Acetic acid phenyl ester

90
59.31

2-Hydroxybenzoic acid (= salicylic acid)

97
60.89

Acetylsalicylic acid

81
62.94

Phenyl salicylate

95
63.84

Xanthone

97
64.79

6H,12H-Dibenzo[b,f][1,5]dioxocin-6,12-dione

(Dimer of 2-Hydroxybenzoic acid)

64
71.02

2,10,18-Trioxatetracyclo[18.4.0.0(4,9).0(12,17)]

tetracosa-1(24),4,6,8,12,14,16,20,22-nonaene-3,11,

19-trione (Trimer of 2-Hydroxybenzoic acid)

90
Measured spectrum comparison of 2-hydroxybenzoic acid (blue) and another sample (red) with labeled peaks for analysis.
3) Measured spectrum at 59.31 min (red, top) compared to the library spectrum of 2-hydroxybenzoic acid (blue, bottom)
Measured mass spectrum comparison of aspirin (blue) and sample data (red) showing key peaks and molecular structure.
4) Measured spectrum at 60.89 min (red, top) compared to the library specrum of aspirin (acetylsalicylic acid, blue, bottom)

Conclusion

The combination of thermogravimetry and GC-MS (gas chromatography / mass spectrometry) is a powerful technique for gaining deep insights into thermal Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition processes and the resulting gases released. The thermal Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition of acetylsalicylic acid in a helium atmosphere results in a complex gas mixture of at least nine different compounds released. Previous studies by TGA-FT-IR (Fourier transform infrared spectroscopy coupled to a thermobalance) showed that the first mass-loss step releases acetic acid and salicylic acid, whereas the second mass-loss step is the result of a complex Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition reaction. The capability of GC-MS starts where FT-IR reaches its limitations and provides much deeper insights into mixtures of simultaneously released gases. TGA-GCMS is capable to both separate and identify them.

Literature

  1. [1]
    www.bayer.com/en/products/aspirin
  2. [2]
    AN 209 – Deeper Insight into the PyrolysisPyrolysis is the thermal decomposition of organic compounds in an inert atmosphere.Pyrolysis of Acetylsalicylic Acid by Means of Kinetic Analysis, Part 2
  3. [3]
    AN 210 – Deeper Insight into the PyrolysisPyrolysis is the thermal decomposition of organic compounds in an inert atmosphere.Pyrolysis of Acetylsalicylic Acid by Means of Thermogravimetric Measurementsin Various Gas Atmopshere, Part 3
  4. [4]
    AN 211 – Deeper Insight into the PyrolysisPyrolysis is the thermal decomposition of organic compounds in an inert atmosphere.Pyrolysis of Acetylsalicylic Acid by Means of Thermogravimetry and Mass Spectrometry, Part 4
  5. [5]
    Gregory T. Long, Sergey Vyazovkin, Nicoleigh Gamble, Charles A. Wight, Journal of Pharmaceutical Sciences, Vol. 91, No. 3, March 2002
  6. [6]
    AN 136 – About the Thermal Behavior of Acetylsalicylic Acid and Aspirin