Veröffentlicht am 

Comparison of a Blue Diode Laser with a Mercury Arc Lamp in the Curing of a Water-Soluble Resin by Photo-DSC and Photo-DEA

Introduction

Photo curing of liquid monomers and oligomers is employed in a variety of industries as an environmentally friendly, safe, fast, and easily controlled approach to forming inks, coatings, adhesives, and structural materials. The expansion of applications for photo curing since its introduction in the 1960s has been accompanied by an evolution in the light sources used. For example, stereolithography, an additive process for manufacturing 3-dimensional objects from photo-curable polymer resin, requires a laser to trace complex patterns on each layer of liquid resin. 

The ability to measure curing kinetics and degree of cure is essential to the selection of suitable UV and visible light sources, the identification of optimal curing times and conditions, and the development of new, photo curable resins. Photo-differential scanning calorimetry (Photo-DSC) and photo-dielectric analysis (Photo- DEA) are powerful analytical tools for accomplishing these measurements. 

In the example presented here, the efficiencies of two different UV light sources were compared in the curing of a water-soluble, blue-curing adhesive. Laser curing was employed, for the first time, in combination with DSC and DEA measurements and compared with the standard mercury (Hg) arc lamp. The pre-polymer formulation consisted of polyethylene glycol diacrylate (PEGDA) with camphorquinone (CQ) photo-initiator (1% by weight relative to PEGDA) and N,N-dimethyl-p-toluidine (DMPT) as a coinitiator (1:1 by weight relative CQ). This formulation has been used to fabricate complex hydrogel scaffolds with a fully interconnected pore network for use as bioreactors1.

1Paul Calvert, Swati MIshra, Amrut Sadacher, Dapeng LI, University of Massachusetts, Dartmouth, NTC Project: F06-MD14, National Textile Center Research Briefs: June 2010

Photo-DSC Measurements

DSC measurements were performed using a NETZSCH DSC 204 F1 Phoenix® interfaced with either an OmniCure® S2000 200 watt Hg short-arc lamp (Figure 1) with a band pass filter delivering a spectral range of 320-500 nm with an irridiance of 10 W/cm² or a LASERGLO W Technologies LRD-0447 Series collimated diode laser system (Figure 2) delivering 447 nm wavelength of 0.744 W/cm2.

NETZSCH DSC 204 F1 Phoenix thermal analysis system with OmniCure S2000, showcasing precise measurement capabilities.
1) NETZSCH DSC 204 F1 Phoenix® with OmniCure® S2000 and light guide
Collimated diode laser system LRD-0447 from LASERGLOW, showcasing a bright blue laser on a precision surface.
2) LASERGLOW Technologies LRD-0447 Series collimated diode laser system

Figure 3 and Figure 4 show the results of three sets of DSC measurements of the resin curing under multiple 2-second pulses from the Hg arc lamp and from the laser, respectively. Degree of cure calculations based on peak areas from the three lamp runs and the three laser runs are listed in Table 1 and Table 2, respectively. The measurements exhibited good reproducibility. 

The total resin curing enthalpy was greater for the laser (129±5 J/g) than for the lamp (91±6 J/g).2 The corrected enthalpy of each peak from the laser runs was, on average, greater than the corresponding peak from the measurements with the lamp. Furthermore, unlike the lamp, the laser continued to generate additional curing enthalpy up to the final pulse in the measurement. The residual peak area at the end of the cure (e.g. pulse No. 15) is attributable to the heating effect of the light source on the sample, which was nine times greater for the lamp than for the laser.

2Total curing enthalpy was calculated by totaling the peak areas and subtracting out the baseline contribution from differential heating of the sample and reference crucibles, which was calculated from the enthalpy of the final pulse in the series. The timing of the Omnicure lamp pulses was controlled by the NETZSCH Proteus® software. The timing of the laser pulses was controlled manually.

DSC measurement results of resin curing with multiple 2s pulses, illustrated in green, red, and blue curves over time in minutes.
3) Results of three separate DSC measurements, shown in different colors, of resin sample curing under multiple 2s pulses by the Hg arc lamp
DSC graph displaying results of resin sample curing under blue diode laser pulses, highlighting thermal transitions in green, red, and blue.
4) Results of three separate DSC measurements, shown in different colors, of resin sample curing under multiple 2s pulses by the blue diode laser

Table 1: Degree of cure calculations (Hg lamp)

 

First Run

Second Run

Third Run

Pulse

No.

Peak

area

(Jg)

Corrected enthalpy

J/g)

Conversion

(%)

Peak

area

(Jg)

Corrected

enthalpy (J/g)

Conversion

(%)

Peak

area

(J/g)

Corrected

enthalpy

(J/g)

Conversion

(%)

171.4734.1940.5172.9137.8740.2971.2238.0840.24
258.3521.0734.9656.7821.7423.1355.1221.9823.23
349.4212.1414.3847.8512.8113.6345.712.5623.23
444.477.198.5242.547.507.9840.887.748.18
541.594.315.1139.774.735.0338.024.885.16
639.932.653.1438.283.243.4536.383.243.42
738.861.581.8737.252.212.3535.182.042.16
838.130.851.0136.421.381.4734.551.411.49
937.910.630.7536.121.081.1532.211.071.13
1037.500.220.2635.800.760.8133.840.700.74
1137.27-0.01-0.0135.520.480.5133.600.460.49
1237.17-0.11-0.1335.140.100.1133.430.290.31
1337.06-0.12-0.1434.95-0.09-0.1033.290.150.16
1437.09-0.19-0.2335.230.190.2033.170.030.03
1537.280.000.0035.040.000.0033.140.000.00
  

Total enthalpy =

84.40 J/g

  

Total enthalpy =

94.00 J/g

  

Total enthalpy =

94.63 J/g

 

Table 2: Degree of cure calculations (laser)

 

First Run

Second Run

Third Run

Pulse

No.

Peak

area

(Jg)

Corrected enthalpy

J/g)

Conversion

(%)

Peak

area

(Jg)

Corrected

enthalpy (J/g)

Conversion

(%)

Peak

area

(J/g)

Corrected

enthalpy

(J/g)

Conversion

(%)

150.7046.0235.4047.7243.1732.5644.4640.1932.47
229.6024.9219.1733.0128.4621.4732.6128.3422.89
321.6716.9913.0922.9118.3613.8520.3516.0812.99
418.3913.7110.5414.9310.387.8315.7911.529.31
513.128.446.4912.828.276.2410.66.335.11
610.255.574.289.835.283.9810.095.814.69
78.673.993.089.935.384.068.5024.233.42
87.382.692.077.773.222.437.9573.692.98
97.202.521.947.392.842.147.0772.812.27
106.311.621.257.312.762.085.9851.721.39
115.681.000.776.131.581.195.4081.140.92
125.991.301.005.671.120.845.7771.511.22
135.590.900.695.540.990.744.440.170.14
145.020.340.265.330.780.594.5210.250.20
154.690.000.004.550.000.004.2690.000.00
  

Total enthalpy =

128.99 J/g

  

Total enthalpy =

132.58 J/g

  

Total enthalpy =

123.79 J/g

 

Photo-DEA Measurements

DEA monitoring of the resin photocuring process at ambient temperature using the two different light sources were performed with a NETZSCH DEA 288 Epsilon instrument (Figure 5). The results are compared in Figure 6. Two measurements were performed with each radiation source in order to demonstrate reproducibility. Both the laser and lamp were run continuously with the exception of a two minute interruption in irradiation from the lamp during one of the runs. Curing progress is indicated by an increase in Ion Viscosity, which levels off as curing completes. The initial slopes of the Ion Viscosity curves are slightly greater for laser-cured samples than for the lamp-cured ones, indicating more efficient curing from the laser. The overall increase in the Ion Viscosity was also slightly greater for the laser-cured samples. DEA measurements are more sensitive to small changes in the degree of cure than DSC measurements. Hence, increases in the Ion Viscosities of the samples due to curing were still measurable after 50 minutes of continuous lamp or laser irradiation. Due to sample heating by the lamp or laser, which causes an increase in ion mobility, sharp steps in the curves are observed as soon as the light source is removed.

NETZSCH DEA 288 Epsilon laboratory equipment setup, including OmniCure S2000, lab furnace, and computer for material testing.
5) NETZSCH DEA 288 Epsilon laboratory version with OmniCure® S2000, light guide, lab furnace and computer
Ion viscosity curves at 10 Hz show measurements for laser (blue) and lamp (green), highlighting distinct behaviors over time.
6) Ion viscosity curves measured at 10 Hz frequency of curable resion irradiated with Hg arc lamp and laser

Summary

In summary, a comparison of the curing enthalpy and curing kinetics of photo-curable resin under irradiation with a Hg arc lamp and a blue diode laser was accomplished using NETZSCH photo-DSC and photo-DEA instrument configurations. DSC measurements showed that the enthalpy of resin curing with the laser was greater than that with the lamp, indicating possibly greater crosslinking of the sample with the laser. This is consistent with the greater absolute change in the Ion Viscosity of the laser cured sample measured by DEA. The DEA measurements also showed that the resin curing rate was slightly greater with the laser than with the lamp. Finally, the DSC measurements indicated greater sample heating by the Hg lamp radiation than by the laser radiation. Sample heating can be an issue in cases where temperature changes during the polymerization lead to polymer shrinkage stress. Overall, the lower intensity, monochromatic blue laser proved to be a more suitable light source for curing this particular resin formulation than the Hg arc lamp with a broad-band filter.

AI Overview
An error occurred. Please try again.