Published on 

Thermogravimetric Analysis of Redox Reaction of CuO and Cu by Means of the H2Secure Box

Hydrogen: A Key Driver in Clean Energy Transitions

Introduction

Hydrogen stands at the forefront of clean energy transitions, driving carbon-free industrial processes and supporting renewable energy integration. Its versatility in production, storage, and utilization highlights its role as a cornerstone of sustainable energy systems. Recent research leveraging advanced thermal analysis techniques has revealed hydrogen’s broad application potential, including its role in production technologies, metallurgical processes, thermochemical energy storage, and innovative reduction/OxidationOxidation can describe different processes in the context of thermal analysis.oxidation cycles. These advancements underline hydrogen’s transformative impact on energy and material science.

One example is the use of thermogravimetric analysis (TGA) to study reduction/OxidationOxidation can describe different processes in the context of thermal analysis.oxidation cycles of metal oxides/metals for carbon-neutral energy applications. Studies [Chen et al., 2024; Cerciello et al., 2024] have shown that repeated reduction/OxidationOxidation can describe different processes in the context of thermal analysis.oxidation cycles with hydrogen in controlled atmospheres can lead to structural changes that affect reactivity. The results of these papers provide insights into structural changes under non-IsothermalTests at controlled and constant temperature are called isothermal.isothermal and IsothermalTests at controlled and constant temperature are called isothermal.isothermal conditions, revealing the effect of temperature and gas composition on reaction kinetics. In the field of thermochemical energy storage, the OxidationOxidation can describe different processes in the context of thermal analysis.oxidation kinetics ofCu2Oto CuO have been analyzed [Jahromy et al., 2019].

Instrumentation

In this application note, we seek to demonstrate the capability of our new developments for the NETZSCH STA 509 series. These are designed to support advanced hydrogen research, helping to investigate kinetic changes during reversible redox reactions. The system is engineered to handle experiments in a 100% hydrogen atmosphere, addressing the challenges of hydrogen’s flammability risks at temperatures up to 1600°C.

A key innovation is the integration of the H₂Secure system into STA devices, ensuring safe operation in up to 100% H₂ atmospheres. It includes a centralized control box for gas regulation, real-timeH2 andO2 monitoring, and a fail-safe mechanism that purges hydrogen out with inert gas in the case of malfunctions. An optimized gas flow path ensures controlled distribution of the gas atmosphere over the sample. An internal pressure sensor enables monitoring of overpressure limits within the furnace and measurement chambers. This capability allows for the detection of accidental leakage formation during experiments, ensuring enhanced safety and system integrity.

Experimental Results and Discussion

The example in this study highlights the reversible redox reaction of copper oxide (CuO) and copper (Cu) under controlled conditions. A series of cycles were performed at 500°C using 100%H2 for reduction and synthetic air (21%O2) for OxidationOxidation can describe different processes in the context of thermal analysis.oxidation.

The main measurement parameters are listed in table 1.

Table 1: Measurement parameters

InstrumentSTA4491
SampleCuO
Sample mass29.975 mg
CrucibleAl2O3 open
FurnaceSiC
Sample carrierTGA Plate P
AccessoriesH₂SecureBox,H2 Generator

Purge 1

H2 (150 ml/min)

Purge 2

Ar (150 ml/min)

Purge 3

Synthetic air (150 ml/min(

Protective

Ar (20 ml/min)

1 Experiments were conducted using the previous version (STA 449) of the STA 509 series instrument, which is fully compatible with the current version and provides comparable accuracy and result quality.

Figure 1 shows the TGA results obtained. The findings demonstrate the system’s reversibility, with gradual kinetic changes observed over successive cycles.

TGA analysis graph showing copper oxide weight changes at 500°C in alternating hydrogen and air atmospheres, highlighting key data points.
1) TGA results on copper oxide powder at 500°C in an alternating hydrogen (100% H2) and synthetic air (21% O2) atmosphere.

These results are discussed in the following steps.

1. Initial Heating:
The sample was heated to 500°C under a protective argon atmosphere (Purge 2 and Protective).

2. Reduction Phase:

3. Transition to OxidationOxidation can describe different processes in the context of thermal analysis.Oxidation:

4. OxidationOxidation can describe different processes in the context of thermal analysis.Oxidation Phase:

Cycles

The results of this experiment highlight the reversible nature of the CuO/Cu redox reaction

CuO +H2 ↔ Cu +H2O

and demonstrate the impact of surface passivation on reaction kinetics, particularly during the OxidationOxidation can describe different processes in the context of thermal analysis.oxidation step. These findings are critical for understanding material behavior under cyclic redox conditions, with implications for catalytic and energy storage applications.

Summary

The NETZSCH STA 509 Jupiter® in combination with theH₂Securebox constitutes a powerful tool for hydrogen research. The system is designed to analyze high-temperature redox reactions under controlled atmospheres, including hydrogen-rich and mixed gases. Its advanced features ensure safety and reliability during experiments while supporting a wide range of applications, includ-ing the study of reduction-oxidation cycles, the optimi-zation of catalytic processes, and the improvement of hydrogen-based technologies in metallurgy and energy storage. By providing precise insights into reaction kinet-ics, Phase TransitionsThe term phase transition (or phase change) is most commonly used to describe transitions between the solid, liquid and gaseous states.phase transitions, and material stability, the STA 509 series enables researchers to enhance efficiency and sus-tainability in industrial and material applications, propel-ling innovation in hydrogen-driven processes.

Literature

  1. [1]
    Chen, R., Hansen, B. B., Lin, W., Wu, H., & Glarborg, P. (2024). Deactivation of iron particles during com-bustion and reduction. Fuel, 378, 132915. https://doi. org/10.1016/j.fuel.2024.132915
  2. [2]
    Jahromy, S. S., Birkelbach, F., Jordan, C., Huber, C., Harasek, M., Werner, A., & Winter, F. (2019). Impact of partial pressure, conversion, and temperature on the oxidation reaction kinetics of Cu2O to CuO in thermo-chemical energy storage. Energies, 12(508). https://doi. org/10.3390/en12030508
  3. [3]
    Cerciello, F., Fabozzi, A., Yannakis, C., Schmitt, S., Narin, O., Scherer, V., & Senneca, O. (2024). Kinetics of iron reduction upon reduction/oxidation cycles. Interna-tional Journal of Hydrogen Energy, 65, 337–347. https://doi.org/10.1016/j.ijhydene.2024.04.008”t”_new
AI Overview
An error occurred. Please try again.