Published on 

Determination of Yield Stress of Complex Fluids by Oscillation Tests on a Rotational Rheometer – Various Gel Samples

Introduction

Many complex fluids, such as network forming polymers, surfactant mesophases and concentrated emulsions do not flow until the applied StressStress is defined as a level of force applied on a sample with a well-defined cross section. (Stress = force/area). Samples having a circular or rectangular cross section can be compressed or stretched. Elastic materials like rubber can be stretched up to 5 to 10 times their original length.stress exceeds a certain critical value, known as the Yield StressYield stress is defined as the stress below which no flow occurs; literally behaves like a weak solid at rest and a liquid when yielded.yield stress. Materials exhibiting this behavior are said to be exhibiting yield flow behavior. The Yield StressYield stress is defined as the stress below which no flow occurs; literally behaves like a weak solid at rest and a liquid when yielded.yield stress is therefore defined as the StressStress is defined as a level of force applied on a sample with a well-defined cross section. (Stress = force/area). Samples having a circular or rectangular cross section can be compressed or stretched. Elastic materials like rubber can be stretched up to 5 to 10 times their original length.stress that must be applied to the sample before it starts to flow. Below the Yield StressYield stress is defined as the stress below which no flow occurs; literally behaves like a weak solid at rest and a liquid when yielded.yield stress the sample will deform elastically (like stretching a spring), above the Yield StressYield stress is defined as the stress below which no flow occurs; literally behaves like a weak solid at rest and a liquid when yielded.yield stress the sample will flow like a liquid.

Most fluids with Yield StressYield stress is defined as the stress below which no flow occurs; literally behaves like a weak solid at rest and a liquid when yielded.yield stress can be considered as a structural skeleton that extends over the entire volume of the system. The strength of the skeleton is governed by the structure of the dispersed phase and its interactions. Normally, the continuous phase is low in viscosity, however high volume fractions of a dispersed phase can increase the viscosity by a thousand times and induce solid-like behavior at rest.

There are various methods for determining the yield StressStress is defined as a level of force applied on a sample with a well-defined cross section. (Stress = force/area). Samples having a circular or rectangular cross section can be compressed or stretched. Elastic materials like rubber can be stretched up to 5 to 10 times their original length.stress [1], mostly using steady shear testing, however, one of the most sensitive methods is using an oscillatory amplitude sweep. This test involves applying an increasing StressStress is defined as a level of force applied on a sample with a well-defined cross section. (Stress = force/area). Samples having a circular or rectangular cross section can be compressed or stretched. Elastic materials like rubber can be stretched up to 5 to 10 times their original length.stress or StrainStrain describes a deformation of a material, which is loaded mechanically by an external force or stress. Rubber compounds show creep properties, if a static load is applied.strain and monitoring changes in modulus and/or StressStress is defined as a level of force applied on a sample with a well-defined cross section. (Stress = force/area). Samples having a circular or rectangular cross section can be compressed or stretched. Elastic materials like rubber can be stretched up to 5 to 10 times their original length.stress. 

There are different ways of interpreting yield StressStress is defined as a level of force applied on a sample with a well-defined cross section. (Stress = force/area). Samples having a circular or rectangular cross section can be compressed or stretched. Elastic materials like rubber can be stretched up to 5 to 10 times their original length.stress from an amplitude sweep as shown in Figure 1. Some workers consider the initial drop in G’ a measure of the yield point since this is represents the onset of non-linearity and structural breakdown, while others consider the G’/G” cross-over to be the yield point since this represents the transition from solid to liquid-like behavior. The zone between these events is often referred to as the yield zone.

Yield StressStress is defined as a level of force applied on a sample with a well-defined cross section. (Stress = force/area). Samples having a circular or rectangular cross section can be compressed or stretched. Elastic materials like rubber can be stretched up to 5 to 10 times their original length.stress is defined as the StressStress is defined as a level of force applied on a sample with a well-defined cross section. (Stress = force/area). Samples having a circular or rectangular cross section can be compressed or stretched. Elastic materials like rubber can be stretched up to 5 to 10 times their original length.stress that must be applied to the sample before it starts to flow.

Diagram illustrating amplitude sweep data with yield point determination points marked as 1 and 2 on a stress vs. modulus graph.
1) Diagram showing the points commonly used from an amplitude sweep to determine the yield point

A more recent method involves measuring the elastic StressStress is defined as a level of force applied on a sample with a well-defined cross section. (Stress = force/area). Samples having a circular or rectangular cross section can be compressed or stretched. Elastic materials like rubber can be stretched up to 5 to 10 times their original length.stress component, σ’ (associated with the elastic structure through G’) as a function of StrainStrain describes a deformation of a material, which is loaded mechanically by an external force or stress. Rubber compounds show creep properties, if a static load is applied.strain amplitude. The yield StressStress is defined as a level of force applied on a sample with a well-defined cross section. (Stress = force/area). Samples having a circular or rectangular cross section can be compressed or stretched. Elastic materials like rubber can be stretched up to 5 to 10 times their original length.stress is taken as the peak stress, and the StrainStrain describes a deformation of a material, which is loaded mechanically by an external force or stress. Rubber compounds show creep properties, if a static load is applied.strain at this point the yield StrainStrain describes a deformation of a material, which is loaded mechanically by an external force or stress. Rubber compounds show creep properties, if a static load is applied.strain (see Figure 2). Generally, this value fits somewhere within the yield zone and has been shown to give a more reliable measurement of the yield stress which correlates well with other methods.

Test frequency can sometimes affect the measured yield stress depending on the RelaxationWhen a constant strain is applied to a rubber compound, the force necessary to maintain that strain is not constant but decreases with time; this behavior is known as stress relaxation. The process responsible for stress relaxation can be physical or chemical, and under normal conditions, both will occur at the same time. relaxation behavior of the material under test. Lower frequencies will give a better indication of material properties at rest but will greatly increase the time of the test. Consequently, values between 0.1 and 10 Hz are commonly employed.

This application note shows methodology and data for a number of gel samples.

Peak elastic stress diagram illustrating yield stress and strain relationship, crucial for material analysis and testing.
2) Diagram showing the peak in elastic stress component as a function of StrainStrain describes a deformation of a material, which is loaded mechanically by an external force or stress. Rubber compounds show creep properties, if a static load is applied.strain amplitude, which enables determination of yield stress and StrainStrain describes a deformation of a material, which is loaded mechanically by an external force or stress. Rubber compounds show creep properties, if a static load is applied.strain

Experimental

Results and Discussion

Figure 3 shows the results from the StrainStrain describes a deformation of a material, which is loaded mechanically by an external force or stress. Rubber compounds show creep properties, if a static load is applied.strain amplitude sweeps for the diffferent samples, and Table 1 shows the corresponding values of yield stress and strain as determined from an automatic peak analysis.

Elastic stress vs. strain amplitude graph displaying hair gel and mannan/xanthan gum sample data in blue curves.
3) Plots of elastic stress (σ’) against strain amplitude (γ) for hair gel ( ); mannan/xanthan gum (Δ); HASE-surfactant samples

Table 1: Values of yield stress and strain determined from analysis of the elastic stress plots

Sample DescriptionYield StrainStrain describes a deformation of a material, which is loaded mechanically by an external force or stress. Rubber compounds show creep properties, if a static load is applied.StrainYield StressYield stress is defined as the stress below which no flow occurs; literally behaves like a weak solid at rest and a liquid when yielded.Yield Stress (Pa)
Hair gel0.86977.3
Xanthan gum/Mannan1.47223.4
HASE-surfactant0.19411.1

The hair gel has the highest yield stress with a measured value of 77 Pa. The gum complex gave a yield stress of 23 Pa while the associative thickener had the lowest value of 11 Pa.

In terms of yield strain, the highest value was measured for the gum complex with a value of 1.5, thus indicating a more ductile structure. The hair gel gave a value of 0.87 and the associative thickener (HASE-surfactant) a value of 0.2 indicating a more brittle structure.

Conclusion

An oscillation amplitude sweep test can be used to determine the yield stress and strain of a material. The preferred test method involves monitoring the elastic stress σ’ as a function of strain amplitude γ with the yield stress inferred to be the measured peak value of σ’. This test has been used to measure the yield stress and strain of a number of aqueous gel systems.

2Please note that a parallel plate geometry can also be used – with this geometry being preferred for dispersions and emulsions with large particle sizes. Such material types may also require the use of serrated or roughened geometries to avoid artefacts relating to slippage at the geometry surface.

Literature

  1. [1]
AI Overview
An error occurred. Please try again.