NETZSCH DMA 303 instrument displayed with a dynamic background, showcasing advanced dynamic-mechanical analysis capabilities.

14.02.2024 by Rüdiger Sehling, Aileen Sammler

Why Is DMA That Important?

Comparing Measurement Results when Measuring Polytetrafluoroethylene (PTFE) by Means of Differential Scanning Calorimetry (DSC) and Dynamic-Mechanical Analysis (DMA)

When measuring polymer materials with a DSC (Differential Scanning Calorimeter), it can be difficult to monitor effects such as the Glass Transition TemperatureThe glass transition is one of the most important properties of amorphous and semi-crystalline materials, e.g., inorganic glasses, amorphous metals, polymers, pharmaceuticals and food ingredients, etc., and describes the temperature region where the mechanical properties of the materials change from hard and brittle to more soft, deformable or rubbery.glass transition. By means of DSC instruments, only the energetic effects of materials (endothermal/ExothermicA sample transition or a reaction is exothermic if heat is generated.exothermal) are measured, i.e., the change in specific heat. With a DMA (Dynamic Mechanical Analyzer), however, it is not possible to detect energetic effects since the real mechanical material behavior is determined, and its change in mechanical properties (especially during the Glass Transition TemperatureThe glass transition is one of the most important properties of amorphous and semi-crystalline materials, e.g., inorganic glasses, amorphous metals, polymers, pharmaceuticals and food ingredients, etc., and describes the temperature region where the mechanical properties of the materials change from hard and brittle to more soft, deformable or rubbery.glass transition) is much more sensitive compared to energetic effects.

Figure 1 shows a typical DSC measurement on PTFE. Only two small endothermal effects can be seen from the change in crystalline structure. No more information can be gathered, although PTFE provides much more information.

DSC measurement graph of PTFE, highlighting endothermal peaks at 21°C and 31°C, indicating thermal properties.
Figure 1: DSC measurement on PTFE




Presented in figure 2 is a direct comparison of the DSC and DMA measurements on PTFE. The red curve shows the DSC results and the black curve the DMA results. The continuous black line represents the storage modulus, E’ (stiffness), and the black dashed curve the loss factor tand (damping). In the DMA measurement, it can clearly be seen that definitely more information is obtained compared to DSC. At the beginning in the low-temperature range a transition represented by the drop in storage modulus, E’, at -124°C (E’ onset) with a corresponding maximum in the loss factor, tand, at -104°C (tand peak) can be observed. This is the β-transition of PTFE. Another transition is found in the storage modulus, E’, at 19°C (E’ onset) which represents the solid/solid transition of PTFE that is also measurable by DSC. This transition is also associated with a peak maximum in the loss factor, tan d, at 29°C (tan d peak).
 

Comparison of DMA and DSC measurements on PTFE, highlighting temperature transitions and mechanical properties of the polymer.
Figure 2: Comparison of DMA and DSC measurements on PTFE



The Glass Transition TemperatureThe glass transition is one of the most important properties of amorphous and semi-crystalline materials, e.g., inorganic glasses, amorphous metals, polymers, pharmaceuticals and food ingredients, etc., and describes the temperature region where the mechanical properties of the materials change from hard and brittle to more soft, deformable or rubbery.glass transition of PTFE can be found at higher temperatures by the drop in storage modulus, E’, at 113°C with a corresponding peak maximum in the loss factor, tan d, at 128°C.

It can clearly be seen that DMA is a very sensitive method to detect Phase TransitionsThe term phase transition (or phase change) is most commonly used to describe transitions between the solid, liquid and gaseous states.phase transitions of materials, which are nearly impossible to detect by means of DSC.

Dynamic Mechanical Analysis (DMA) illustration with wave patterns, highlighting material behavior graphs for viscoelastic properties.
Have you already heard of the new NETZSCH DMA 303 Eplexor®? 
In this new single tabletop DMA, we combine highest force with widest temperature range.

Our latest development, the new NETZSCH DMA 303 Eplexor®, is designed for precise measurements on a broad variety of specimens including even very stiff samples with a controlled force range of up to 50 N, both statically and dynamically. The full resolution is available throughout the entire force range, resulting in accurate and reliable data.

The temperature-controlled furnace features an unprecedented broad temperature range of -170°C to 800°C, allowing for homogenous heat distribution around the sample. In addition, the force displacement range of ±30 mm is perfect for static experiments including creep and relaxation.

Advanced NETZSCH DMA 303 analyzer against a sleek, digital backdrop, showcasing precision in dynamic-mechanical analysis for polymers.

Gain Valuable Information with Dynamic-Mechanical Analysis

Dynamic Mechanical Analysis offers a wealth of information about your material:

For more information visit our product page: