TG-GC-MS Measurements on Acetylsalicylic Acid (ASA)

TG 209 F1 Libra thermogravimetric analyzer paired with GC-MS for precise analytical testing and material analysis.
1) TG 209 F1 Libra® with GC-MS coupling

Measurement Parameters

TG 209 F1 Libra® - GC-MS 

TGA Parameters: 

  • RT to 500°C
  • 10 K/min
  • 50 ml/min N
  • Sample mass: 1.71 mg
  • Al2O3 crucibles

 

GC Parameters: 

  • Quasi-continuous
  • GC furnace: 250°C
  • Split: 10:1
  • Column: HP 5 ms, 30 m
  • Valve switching: 30 sec

Introduction

ASA is one of the most widely used non-opioid analgesics. By esterification of the phenolic hydroxyl group of salicylic acid with acetic acid, not only is a better local tolerance achieved, but also a stronger antipyretic, antiphlogistic and especially platelet aggregation inhibiting effect [1]. 

The common name is 2-acetoxybenzoic acid. It forms white, needle-shaped crystals which smell faintly of acetic acid. ASA is produced by means of the Kolbe-Schmitt synthesis, an electrophilic aromatic substitution [2].

[1] Mutschler, Arzneimittelwirkungen

[2] Laue, Reaktionsmechanismen

Chemical structure of acetylsalicylic acid (aspirin) highlighting its molecular connections and composition.
2) Chemical structure of acetylsalicylic acid
Thermogravimetric analysis results with TG (black), DTA (green), c-DTA (red), and total ion chromatogram (purple) for BCR-276.
3) TGA results (black), DTA results (green), c-DTA® signal (red) and total ion chromatogram of the ASA measurements (purple)

Measurement Results

Simultaneous coupling of thermogravimetric analysis (TGA) and GC-MS (gas chromatography and mass spectrometry) enables easy analysis of the thermal behavior and Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition products. 

The melting of ASA can be detected by means of c-DTA® at 142°C. ASA decomposes in two steps (DTG minima at 175°C and 350°C). The total ion chromatogram reflects the completely measured spectra as a function of temperature. Corresponding to DTG, two maxima can clearly be seen. 

The 1st TG step can be attributed to the formation of acetic acid (mass numbers 43 and 60 figure 4); the 2nd step can be attributed to phenol (mass number 94) as a Decomposition reactionA decomposition reaction is a thermally induced reaction of a chemical compound forming solid and/or gaseous products. decomposition product.

TGA results display the thermal analysis curve with total ion chromatograms for mass numbers 43, 60, and 94, indicating thermal behavior.
4) TGA results (black) and total ion chromatogram of mass numbers 43, 60 and 94